Comparison of four different pediatric 10F aortic cannulae during pulsatile versus nonpulsatile perfusion in a simulated neonatal model of cardiopulmonary bypass
- PMID: 18043165
- DOI: 10.1097/MAT.0b013e31815b0cd7
Comparison of four different pediatric 10F aortic cannulae during pulsatile versus nonpulsatile perfusion in a simulated neonatal model of cardiopulmonary bypass
Abstract
We compared four commercially available 10F pediatric aortic cannulae with different geometric designs (DLP-Long tip, DLP-Short tip, RMI-Long tip, and Surgimedics-Short tip) during pulsatile versus nonpulsatile perfusion in terms of pressure drops and surplus hemodynamic energy (SHE) levels in an in vitro neonatal model of cardiopulmonary bypass. The pseudo patient was subjected to seven pump flow rates at 100 ml/min increments in the 400-1,000 ml/min range. A total of 44 experiments (n = 22, nonpulsatile; n = 22, pulsatile) were performed at each of the seven flow rates. Surgimedics had significantly higher pressure drops than the other three cannulae at various flow rates during nonpulsatile and pulsatile perfusion, respectively. When the perfusion mode was changed from nonpulsatile to pulsatile flow, SHE levels at both precannula and postcannula sites increased seven to nine times at all flow rates in all four cannulae. Surgimedics generated a significant lower SHE level when compared with the other three cannulae at all flow rates at both precannula and postcannula sites. The results suggest that different geometries of aortic cannulae have a significant impact on pressure drops of the cannulae as well as hemodynamic energy generation and delivery. Pulsatile perfusion generates more "extra" hemodynamic energy when compared with the nonpulsatile perfusion mode with all four cannulae used in this study.
Similar articles
-
A performance evaluation of eight geometrically different 10 Fr pediatric arterial cannulae under pulsatile and nonpulsatile perfusion conditions in an infant cardiopulmonary bypass model.ASAIO J. 2008 May-Jun;54(3):306-15. doi: 10.1097/MAT.0b013e3181744071. ASAIO J. 2008. PMID: 18496282
-
Quantification of perfusion modes in terms of surplus hemodynamic energy levels in a simulated pediatric CPB model.ASAIO J. 2006 Nov-Dec;52(6):712-7. doi: 10.1097/01.mat.0000249013.15237.5e. ASAIO J. 2006. PMID: 17117064
-
A hemodynamic evaluation of the Medos Deltastream DP1 rotary pump and Jostra HL-20 roller pump under pulsatile and nonpulsatile perfusion in an infant cardiopulmonary bypass model--a pilot study.ASAIO J. 2008 Sep-Oct;54(5):529-33. doi: 10.1097/MAT.0b013e318186bd8d. ASAIO J. 2008. PMID: 18812747
-
Pulsatile perfusion during cardiopulmonary bypass procedures in neonates, infants, and small children.ASAIO J. 2007 Nov-Dec;53(6):706-9. doi: 10.1097/MAT.0b013e318158e3f9. ASAIO J. 2007. PMID: 18043152 Review.
-
An evaluation of the benefits of pulsatile versus nonpulsatile perfusion during cardiopulmonary bypass procedures in pediatric and adult cardiac patients.ASAIO J. 2006 Jul-Aug;52(4):357-61. doi: 10.1097/01.mat.0000225266.80021.9b. ASAIO J. 2006. PMID: 16883112 Review.
Cited by
-
International conference on pediatric mechanical circulatory support systems and pediatric cardiopulmonary perfusion: outcomes and future directions.ASAIO J. 2008 Mar-Apr;54(2):141-6. doi: 10.1097/MAT.0b013e318167afdd. ASAIO J. 2008. PMID: 18356645 Free PMC article. No abstract available.
-
An in vitro fluid dynamic study of pediatric cannulae: the value of animal studies to predict human flow.J Biomech Eng. 2012 Apr;134(4):044501. doi: 10.1115/1.4006428. J Biomech Eng. 2012. PMID: 22667684 Free PMC article.
-
Performance Evaluation of Geometrically Different Pediatric Arterial Cannulae in a Pediatric Cardiopulmonary Bypass Model.Braz J Cardiovasc Surg. 2023 Nov 9;39(1):e20230110. doi: 10.21470/1678-9741-2023-0110. Braz J Cardiovasc Surg. 2023. PMID: 37947184 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources