Comparison of corneal flap thickness between primary eyes and fellow eyes using the Zyoptix XP microkeratome

Affiliations
Comparative Study

Comparison of corneal flap thickness between primary eyes and fellow eyes using the Zyoptix XP microkeratome

Thomas Ho et al. J Cataract Refract Surg. 2007 Dec.

Abstract

Purpose: To evaluate corneal flap thickness created in myopic laser in situ keratomileusis (LASIK) using the Zyoptix XP 120 microkeratome (Bausch & Lomb).

Setting: University-based eye clinic.

Methods: Corneal thickness was measured preoperatively and intraoperatively after flap creation in 62 consecutive patients (124 eyes) who had LASIK for the correction of myopia. Corneal flap thickness was calculated by subtracting stromal bed thickness from total corneal thickness. In each patient, both corneas (right followed by left) were cut by 1 the same Zyoptix XP 120 microkeratome blade at the same session.

Results: The mean actual flap thickness was 115.34 microm +/- 16.34 (SD) in right eyes and 104.55 +/- 14.34 mum in left eyes. The mean actual flap thickness in right eyes was not statistically significantly different from the 120 microm proposed by the manufacturer (P = .142); however, the mean actual flap thickness in left eyes was statistically significantly different from the 120 microm (P<.001). The mean flap thickness in the second eye was also statistically significantly thinner than in the first eye (P<.001).

Conclusions: The corneal flaps were thinner than expected when a Zyoptix XP 120 microkeratome was used in LASIK. Although the first cuts produced thinner flaps, this difference was not statistically significant. However, the second cuts produced significantly thinner flaps.