Image quality with multifocal intraocular lenses and the effect of pupil size: comparison of refractive and hybrid refractive-diffractive designs
- PMID: 18053913
- DOI: 10.1016/j.jcrs.2007.07.035
Image quality with multifocal intraocular lenses and the effect of pupil size: comparison of refractive and hybrid refractive-diffractive designs
Abstract
Purpose: To analyze the image quality with a refractive and 2 hybrid refractive-diffractive multifocal intraocular lenses (IOLs) in vitro to determine the IOL modulation transfer function (MTF) following the EN-ISO international standard for distance and near vision and to study the IOLs' behavior with different pupil sizes.
Setting: Fundación Oftalmológica del Mediterráneo, Valencia, Spain.
Methods: This study evaluated 4 IOLs: the hybrid refractive-diffractive AcrySof ReSTOR SN60D3 (Alcon) and Tecnis ZM900 (AMO), the refractive ReZoom NXG (AMO), and the monofocal AcrySof SN60WF (Alcon). The MTF was calculated from the cross-line spread function recorded with the OPAL Vector System (Image Science Ltd.) using fast Fourier transform techniques. The artificial eye model simulated in vivo conditions of the anterior chamber and included an artificial cornea and physiological solution, in which the IOLs were positioned. The pupil sizes varied between 2.0 mm and 5.0 mm in steps of 0.5 mm.
Results: For distance vision and small pupils (2.0 to 3.5 mm), the refractive IOL provided better image quality than the 2 hybrid IOLs. When the pupil was larger, the distance vision was similar with all IOLs. The 2 hybrid IOLs gave better image quality than the refractive IOL for near vision with any pupil size. With the refractive IOL and pupils smaller than 3.5 mm, near focusing did not occur, resulting in poor image quality. Of the hybrid IOLs, the AcrySof ReSTOR SN60D3 provided better distance image quality and the Tecnis ZM900 provided better near image quality.
Conclusions: The refractive IOL gave better image quality than the hybrid IOLs at distance and with small pupils. With pupils larger than 3.5 mm, the quality was similar. Hybrid IOLs gave significantly better image quality for near vision with all pupil sizes. AcrySof ReSTOR SN60D3 IOLs give better distance vision than Tecnis ZM900 IOLs; the latter gave better near vision. The reference monofocal IOL provided better distance images than any multifocal IOL with all pupil sizes.
Comment in
-
ISO eye model not valid for assessing aspherical lenses.J Cataract Refract Surg. 2008 Jul;34(7):1056-7. doi: 10.1016/j.jcrs.2008.03.038. J Cataract Refract Surg. 2008. PMID: 18571061 No abstract available.
Similar articles
-
One-year outcomes with new-generation multifocal intraocular lenses.Ophthalmology. 2008 Sep;115(9):1508-16. doi: 10.1016/j.ophtha.2008.04.017. Epub 2008 Jun 5. Ophthalmology. 2008. PMID: 18538402 Clinical Trial.
-
Optical performance of monofocal and multifocal intraocular lenses in the human eye.J Cataract Refract Surg. 2008 May;34(5):755-62. doi: 10.1016/j.jcrs.2007.12.038. J Cataract Refract Surg. 2008. PMID: 18471629
-
Modulation transfer function and pupil size in multifocal and monofocal intraocular lenses in vitro.J Cataract Refract Surg. 2005 Dec;31(12):2379-85. doi: 10.1016/j.jcrs.2005.10.024. J Cataract Refract Surg. 2005. PMID: 16473235
-
Vision with bifocal intraocular lenses.Ger J Ophthalmol. 1993 Feb;2(1):32-41. Ger J Ophthalmol. 1993. PMID: 8435655 Review.
-
Multifocal intraocular lenses.Ophthalmol Clin North Am. 2006 Mar;19(1):89-105, vi. doi: 10.1016/j.ohc.2005.09.002. Ophthalmol Clin North Am. 2006. PMID: 16500531 Review.
Cited by
-
Effect of pupil size on optical quality parameters in astigmatic eyes using a double-pass instrument.Biomed Res Int. 2013;2013:124327. doi: 10.1155/2013/124327. Epub 2013 Jun 23. Biomed Res Int. 2013. PMID: 23865038 Free PMC article.
-
Comparison of three different diffractıve multifocal intraocular lenses with a +2.5, +3.0, and +3.75 diopter additıon power.Saudi J Ophthalmol. 2019 Oct-Dec;33(4):353-362. doi: 10.1016/j.sjopt.2019.09.007. Epub 2019 Nov 6. Saudi J Ophthalmol. 2019. PMID: 31933530 Free PMC article.
-
Ray propagation imaging and optical quality evaluation of different intraocular lens models.PLoS One. 2020 Feb 4;15(2):e0228342. doi: 10.1371/journal.pone.0228342. eCollection 2020. PLoS One. 2020. PMID: 32017784 Free PMC article.
-
In vitro optical quality measurements of three intraocular lens models having identical platform.BMC Ophthalmol. 2017 Jun 29;17(1):108. doi: 10.1186/s12886-017-0460-0. BMC Ophthalmol. 2017. PMID: 28662629 Free PMC article.
-
Opto-Mechanical Eye Models, a Review on Human Vision Applications and Perspectives for Use in Industry.Sensors (Basel). 2022 Oct 10;22(19):7686. doi: 10.3390/s22197686. Sensors (Basel). 2022. PMID: 36236784 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials
Miscellaneous