Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2008 Feb;145(2):228-232.
doi: 10.1016/j.ajo.2007.09.030. Epub 2007 Dec 11.

Comparison of central corneal thickness using anterior segment optical coherence tomography vs ultrasound pachymetry

Affiliations
Comparative Study

Comparison of central corneal thickness using anterior segment optical coherence tomography vs ultrasound pachymetry

Hanna Y Kim et al. Am J Ophthalmol. 2008 Feb.

Abstract

Purpose: To determine if there is a systematic difference in central corneal thickness (CCT) measured using anterior segment optical coherence tomography (AS-OCT) as compared with ultrasound pachymetry.

Design: Prospective observational study.

Methods: Consecutive subjects with clinically normal corneas underwent CCT measurement by both ultrasound and AS-OCT while participating in a population-based study in Ghana, West Africa. One eye of each subject was randomly selected for analysis. Two measurements were taken and averaged. Agreement and interobserver variability were also analyzed.

Results: One hundred and fifty-five subjects of African ethnicity and average age 57 years (standard deviation [SD] 12; range, 40 to 98 years) were included. Measurements by AS-OCT and US were taken a mean of 15 days (maximum, six weeks) apart. The mean (SD) [range] US CCT was 525.3 microm (33.5) [422, 653] and 499.0 microm (32.0) [428, 613] with AS-OCT. Measurements by the two modalities were strongly correlated (r(2) = 0.82; P < .001), and a significant difference was observed between mean US and AS-OCT CCT (SD) [range] of 26.3 microm (14.2) [-63, 12] (P < .001). The width of the limits of agreement was 28 microm, about 6% of the average pachymetry reading. In 50 eyes randomly remeasured with OCT by a second observer, the intraclass correlation coefficient was 0.91. There was a small but significant systematic difference between observers (mean 6.9 microm, SD 10.9 microm), or 1.4% (P < .001), increasing the difference noted above.

Conclusion: There is a reproducible systematic difference between CCT measurements taken with ultrasound and OCT. It is important to note in clinical practice, that measurements acquired by these two modalities are not directly interchangeable.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Cross-sectional image of a subject’s anterior chamber acquired by AS-OCT, demonstrating the method used to measure CCT. The corneal apex was identified after correcting ocular rotation, from the peak of the reflectivity profile on the horizontal axis (below the image). The callipers were then aligned on the peak reflections of the anterior and posterior tissue boundaries of the cornea in the axis of the corneal apex (to the right of the image). The two measurements were averaged for each eye.
Figure 2
Figure 2
Scatterplot of CCT measurements by AS-OCT versus ultrasonography for each subject. Almost all measurements lie below the equivalence line demonstrating that CCT measurements are slightly higher with ultrasonography than AS-OCT in almost all subjects.
Figure 3
Figure 3
Bland-Altman plot of AS-OCT – ultrasonography difference versus average of ultrasonic and AS-OCT CCT measurements. This demonstrates the level of the mean difference between the two instruments (bold horizontal line), the limits of agreement (horizontal lines above and below the bold line), and that the systematic difference observed is unrelated to mean CCT.
Figure 4
Figure 4
Scatterplot of AS-OCT pachymetry measurements made by two observers (N=50), demonstrating high correlation (r2=0.91) but with observer 1 measuring slightly higher CCT readings (mean difference 6.9 µm, P < 0.001) than observer 2 in most subjects.

References

    1. Wolffsohn JS, Davies LN. Advances in anterior segment imaging. Curr Opin Ophthalmol. 2007;18:32–38. - PubMed
    1. Doughty MJ, Zaman ML. Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach. Surv Ophthalmol. 2000;44:367–408. - PubMed
    1. Wirbelauer C, Scholz C, Hoerauf H, et al. Noncontact corneal pachymetry with slit lamp-adapted optical coherence tomography. Am J Ophthalmol. 2002;133:444–450. - PubMed
    1. Wong AC, Wong CC, Yuen NS, Hui SP. Correlational study of central corneal thickness measurements on Hong Kong Chinese using optical coherence tomography, Orbscan and ultrasound pachymetry. Eye. 2002;16:715–721. - PubMed
    1. Bechmann M, Thiel MJ, Roesen B, et al. Central corneal thickness determined with optical coherence tomography in various types of glaucoma. Br J Ophthalmol. 2000;84:1233–1237. - PMC - PubMed

Publication types