Clinical performance of highly cross-linked polyethylenes in total hip arthroplasty
- PMID: 18056513
- DOI: 10.2106/JBJS.G.00043
Clinical performance of highly cross-linked polyethylenes in total hip arthroplasty
Erratum in
- J Bone Joint Surg Am. 2008 Mar;90(3):628
Abstract
Aseptic loosening secondary to wear-debris-induced osteolysis has been identified as the leading cause of late failure of total hip arthroplasty. Highly cross-linked polyethylene acetabular liners were developed as one approach to reducing this wear. Preclinical laboratory wear testing showed a number of cross-linked polyethylenes to have dramatically less wear than the polyethylene that had been in use for several decades. After the initial bedding-in phase (one to two years), the percent reductions in the wear rate, as indicated by the amount of penetration of the head into the socket evident on serial radiographs, have been comparable with what was predicted from preclinical hip-simulator testing of the highly cross-linked polyethylenes. To our knowledge, there have been no reports of clinically relevant osteolysis that was clearly attributable to wear of a highly cross-linked polyethylene acetabular liner. However, the clinical performance of these materials should be closely monitored with long-term follow-up.
Similar articles
-
Radiostereometric analysis study of tantalum compared with titanium acetabular cups and highly cross-linked compared with conventional liners in young patients undergoing total hip replacement.J Bone Joint Surg Am. 2015 Apr 15;97(8):627-34. doi: 10.2106/JBJS.N.00605. J Bone Joint Surg Am. 2015. PMID: 25878306 Free PMC article. Clinical Trial.
-
The seven-year wear of highly cross-linked polyethylene in total hip arthroplasty: a double-blind, randomized controlled trial using radiostereometric analysis.J Bone Joint Surg Am. 2011 Apr 20;93(8):716-22. doi: 10.2106/JBJS.J.00287. J Bone Joint Surg Am. 2011. PMID: 21508278 Clinical Trial.
-
[Research progress of backside wear in acetabular liners].Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2013 Dec;27(12):1453-6. Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2013. PMID: 24640364 Review. Chinese.
-
Crosslinked polyethylene compared to conventional polyethylene in total hip replacement: pre-clinical evaluation, in-vitro testing and prospective clinical follow-up study.Acta Orthop. 2006 Oct;77(5):719-25. doi: 10.1080/17453670610012890. Acta Orthop. 2006. PMID: 17068701 Clinical Trial.
-
Wear products of total hip arthroplasty: The case of polyethylene.Morphologie. 2017 Mar;101(332):1-8. doi: 10.1016/j.morpho.2016.06.001. Epub 2016 Jul 15. Morphologie. 2017. PMID: 27426252 Review.
Cited by
-
Radiostereometric analysis study of tantalum compared with titanium acetabular cups and highly cross-linked compared with conventional liners in young patients undergoing total hip replacement.J Bone Joint Surg Am. 2015 Apr 15;97(8):627-34. doi: 10.2106/JBJS.N.00605. J Bone Joint Surg Am. 2015. PMID: 25878306 Free PMC article. Clinical Trial.
-
Successful long-term fixation and progression of osteolysis associated with first-generation cementless acetabular components retrieved post mortem.J Bone Joint Surg Am. 2012 Oct 17;94(20):1877-85. doi: 10.2106/JBJS.J.01507. J Bone Joint Surg Am. 2012. PMID: 23079880 Free PMC article.
-
Total Hip Arthroplasty Using Metal Head on a Highly Cross-linked Polyethylene Liner.Hip Pelvis. 2015 Dec;27(4):216-22. doi: 10.5371/hp.2015.27.4.216. Epub 2015 Dec 30. Hip Pelvis. 2015. PMID: 27536629 Free PMC article.
-
Polyethylene in knee arthroplasty: A review.J Clin Orthop Trauma. 2015 Jun;6(2):108-12. doi: 10.1016/j.jcot.2015.01.096. Epub 2015 Feb 19. J Clin Orthop Trauma. 2015. PMID: 25983517 Free PMC article. Review.
-
Use of morphometry to quantify osteolysis after total hip arthroplasty.Clin Orthop Relat Res. 2010 Nov;468(11):3077-83. doi: 10.1007/s11999-010-1377-6. Epub 2010 May 12. Clin Orthop Relat Res. 2010. PMID: 20461482 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources