Usefulness of ambulatory ECG in the diagnosis of sleep-related breathing disorders
- PMID: 18058583
Usefulness of ambulatory ECG in the diagnosis of sleep-related breathing disorders
Abstract
Background: Sleep-related breathing disorders (SRBD) are the additional factor related to poor prognosis in patients with cardiovascular disorders. The apnoea/hypopnoea index (AHI), describing the number of apnoea and hypopnoea episodes per one hour of sleep, has been used as a marker of severity of the disorder. The disease is present in 4% of men and 2% of women above 40 years of age. However, SRBD are diagnosed in less than 3% of patients with this syndrome due to lack of awareness of the disease among health care practitioners and patients. Polysomnography (PSG) has been used as a golden standard for detecting SRBD, however this test is available only in selected centres. Therefore, a simple, fast and inexpensive test for screening for SRBD is necessary. Respiratory activity influences the amplitude of ECG signal whereas heart rate variability (HRV) depicts the activity of the autonomic nervous system. These associations have been used to develop a new method for detection of SRBD involving analysis of HRV and morphology of ECG signal in ECG monitoring.
Aim: Assessment of accuracy of SRBD detection using estimated AHI (Est.AHI), calculated from Holter ECG recordings.
Methods: In a study group consisting of 74 patients tested for SRBD, simultaneous PSG and 24-hour ECG monitoring were performed. Following PSG, AHI for each patient was calculated. According to the AHI values patients were classified as SRBD patients (AHI >15), non-SRBD patients (AHI <5), whereas 12 individuals had borderline SRBD (5< or = AHI < or =15). Age, prevalence of concomitant disorders and treatment were similar in all groups. In all individuals the Est.AHI value was calculated based on ECG recording. Considering the AHI value as a reference parameter discriminating SDB and non-SRBD patients, the number of false positive and false negative results for detecting SDB with the Est.AHI was calculated. Moreover, the SRBD detection accuracy using the Est.AHI calculation was evaluated by the receiver-operator characteristic (ROC) curves which were used to calculate area under curve (AUC), sensitivity, specificity, as well as positive (PPV) and negative (NPV) predictive values for optimal cut-off value.
Results: According to Est.AHI, 50 (68%) patients were correctly diagnosed. The ROC analysis showed high accuracy of SRBD detection using Est.AHI: AUC - 0.91 with sensitivity - 91.2%, specificity - 87.5%, PPV - 88.6%, and NPV - 88.9%. The cut-off value of Est.AHI set at 17 was optimal for the differentiation between patients with or without SRBD.
Conclusions: The Est.AHI calculated with the Lifescreen Apnea software from Holter ECG is an accurate, specific and sensitive method for the detection and classification of obstructive and mixed SRBD.
Similar articles
-
Sleep apnea diagnosis using an ECG Holter device including a nasal pressure (NP) recording: validation of visual and automatic analysis of nasal pressure versus full polysomnography.Sleep Med. 2009 Jun;10(6):651-6. doi: 10.1016/j.sleep.2008.07.002. Epub 2008 Nov 22. Sleep Med. 2009. PMID: 19028140
-
Screening for sleep-related breathing disorders by transthoracic impedance recording integrated into a Holter ECG system.J Sleep Res. 2006 Dec;15(4):455-62. doi: 10.1111/j.1365-2869.2006.00554.x. J Sleep Res. 2006. PMID: 17118103
-
Heart rate increment: an electrocardiological approach for the early detection of obstructive sleep apnoea/hypopnoea syndrome.Clin Sci (Lond). 2004 Jul;107(1):105-10. doi: 10.1042/CS20040036. Clin Sci (Lond). 2004. PMID: 14992679
-
The scoring of respiratory events in sleep: reliability and validity.J Clin Sleep Med. 2007 Mar 15;3(2):169-200. J Clin Sleep Med. 2007. PMID: 17557426 Review.
-
[Diagnosis of sleep apnea: initial experiences with a staged procedure].Pneumologie. 1989 Nov;43 Suppl 1:587-90. Pneumologie. 1989. PMID: 2692020 Review. German.
Cited by
-
Remote Monitoring of Sympathovagal Imbalance During Sleep and Its Implications in Cardiovascular Risk Assessment: A Systematic Review.Bioengineering (Basel). 2024 Oct 19;11(10):1045. doi: 10.3390/bioengineering11101045. Bioengineering (Basel). 2024. PMID: 39451420 Free PMC article.
-
Flow-mediated skin fluorescence: A novel method for the estimation of sleep apnea risk in healthy persons and cardiac patients.Cardiol J. 2022;29(6):948-953. doi: 10.5603/CJ.a2020.0139. Epub 2020 Nov 3. Cardiol J. 2022. PMID: 33140392 Free PMC article.
-
Usefulness of extended holter ECG monitoring for serious arrhythmia detection in patients with heart failure and sleep apnea.Ann Noninvasive Electrocardiol. 2013 Mar;18(2):163-9. doi: 10.1111/anec.12012. Epub 2012 Nov 22. Ann Noninvasive Electrocardiol. 2013. PMID: 23530487 Free PMC article.
-
Repeatability of sleep apnea detection in 48-hour holter ECG monitoring.Ann Noninvasive Electrocardiol. 2010 Jul;15(3):218-22. doi: 10.1111/j.1542-474X.2010.00367.x. Ann Noninvasive Electrocardiol. 2010. PMID: 20645963 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Medical
Research Materials