Dual-energy approach to contrast-enhanced mammography using the balanced filter method: spectral optimization and preliminary phantom measurement
Abstract
Dual-energy contrast agent-enhanced mammography is a technique of demonstrating breast cancers obscured by a cluttered background resulting from the contrast between soft tissues in the breast. The technique has usually been implemented by exploiting two exposures to different x-ray tube voltages. In this article, another dual-energy approach using the balanced filter method without switching the tube voltages is described. For the spectral optimization of dual-energy mammography using the balanced filters, we applied a theoretical framework reported by Lemacks et al. [Med. Phys. 29, 1739-1751 (2002)] to calculate the signal-to-noise ratio (SNR) in an iodinated contrast agent subtraction image. This permits the selection of beam parameters such as tube voltage and balanced filter material, and the optimization of the latter's thickness with respect to some critical quantity-in this case, mean glandular dose. For an imaging system with a 0.1 mm thick CsI:T1 scintillator, we predict that the optimal tube voltage would be 45 kVp for a tungsten anode using zirconium, iodine, and neodymium balanced filters. A mean glandular dose of 1.0 mGy is required to obtain an SNR of 5 in order to detect 1.0 mg/cm2 iodine in the resulting clutter-free image of a 5 cm thick breast composed of 50% adipose and 50% glandular tissue. In addition to spectral optimization, we carried out phantom measurements to demonstrate the present dual-energy approach for obtaining a clutter-free image, which preferentially shows iodine, of a breast phantom comprising three major components-acrylic spheres, olive oil, and an iodinated contrast agent. The detection of iodine details on the cluttered background originating from the contrast between acrylic spheres and olive oil is analogous to the task of distinguishing contrast agents in a mixture of glandular and adipose tissues.
Similar articles
-
Phantom study to evaluate contrast-medium-enhanced digital subtraction mammography with a full-field indirect-detection system.Med Phys. 2010 Feb;37(2):577-89. doi: 10.1118/1.3276733. Med Phys. 2010. PMID: 20229866
-
Analytical optimization of digital subtraction mammography with contrast medium using a commercial unit.Med Phys. 2008 Dec;35(12):5544-57. doi: 10.1118/1.3003063. Med Phys. 2008. PMID: 19175112
-
Optimization of contrast-enhanced breast imaging: Analysis using a cascaded linear system model.Med Phys. 2017 Jan;44(1):43-56. doi: 10.1002/mp.12004. Epub 2017 Jan 3. Med Phys. 2017. PMID: 28044312
-
Contrast-enhanced dual-energy mammography: a promising new imaging tool in breast cancer detection.Womens Health (Lond). 2014 May;10(3):289-98. doi: 10.2217/whe.14.18. Womens Health (Lond). 2014. PMID: 24956295 Review.
-
Monochromatic X-rays: The future of breast imaging.Eur J Radiol. 2021 Nov;144:109961. doi: 10.1016/j.ejrad.2021.109961. Epub 2021 Sep 20. Eur J Radiol. 2021. PMID: 34562745 Review.
Cited by
-
In-line spectroscopy for iodine quantification in dynamic contrast-enhanced dedicated breast CT.Med Phys. 2025 Feb;52(2):1037-1044. doi: 10.1002/mp.17497. Epub 2024 Oct 29. Med Phys. 2025. PMID: 39470292 Free PMC article.
-
Pulse pileup analysis for a double-sided silicon strip detector using variable pulse shapes.IEEE Trans Nucl Sci. 2019 Jun;66(6):960-968. doi: 10.1109/TNS.2019.2917144. Epub 2019 May 15. IEEE Trans Nucl Sci. 2019. PMID: 31327872 Free PMC article.
-
Direct measurement of mammographic X-ray spectra with a digital CdTe detection system.Sensors (Basel). 2012;12(6):8390-404. doi: 10.3390/s120608390. Epub 2012 Jun 18. Sensors (Basel). 2012. PMID: 22969406 Free PMC article.
-
Dosimetric characterization of a dedicated breast computed tomography clinical prototype.Med Phys. 2010 Aug;37(8):4110-20. doi: 10.1118/1.3457331. Med Phys. 2010. PMID: 20879571 Free PMC article.
-
Quantification of breast density with dual energy mammography: an experimental feasibility study.Med Phys. 2010 Feb;37(2):793-801. doi: 10.1118/1.3284975. Med Phys. 2010. PMID: 20229889 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical