Quantification of cerebral blood flow and oxygen metabolism with 3-dimensional PET and 15O: validation by comparison with 2-dimensional PET
- PMID: 18077532
- DOI: 10.2967/jnumed.107.044008
Quantification of cerebral blood flow and oxygen metabolism with 3-dimensional PET and 15O: validation by comparison with 2-dimensional PET
Abstract
Quantitative PET with (15)O provides absolute values for cerebral blood flow (CBF), cerebral blood volume (CBV), cerebral metabolic rate of oxygen (CMRO(2)), and oxygen extraction fraction (OEF), which are used for assessment of brain pathophysiology. Absolute quantification relies on physically accurate measurement, which, thus far, has been achieved by 2-dimensional PET (2D PET), the current gold standard for measurement of CBF and oxygen metabolism. We investigated whether quantitative (15)O study with 3-dimensional PET (3D PET) shows the same degree of accuracy as 2D PET.
Methods: 2D PET and 3D PET measurements were obtained on the same day on 8 healthy men (age, 21-24 y). 2D PET was performed using a PET scanner with bismuth germanate (BGO) detectors and a 150-mm axial field of view (FOV). For 3D PET, a 3D-only tomograph with gadolinium oxyorthosilicate (GSO) detectors and a 156-mm axial FOV was used. A hybrid scatter-correction method based on acquisition in the dual-energy window (hybrid dual-energy window [HDE] method) was applied in the 3D PET study. Each PET study included 3 sequential PET scans for C(15)O, (15)O(2), and H(2)(15)O (3-step method). The inhaled (or injected) dose for 3D PET was approximately one fourth of that for 2D PET.
Results: In the 2D PET study, average gray matter values (mean +/- SD) of CBF, CBV, CMRO(2), and OEF were 53 +/- 12 (mL/100 mL/min), 3.6 +/- 0.3 (mL/100 mL), 3.5 +/- 0.5 (mL/100 mL/min), and 0.35 +/- 0.06, respectively. In the 3D PET study, scatter correction strongly affected the results. Without scatter correction, average values were 44 +/- 6 (mL/100 mL/min), 5.2 +/- 0.6 (mL/100 mL), 3.3 +/- 0.4 (mL/100 mL/min), and 0.39 +/- 0.05, respectively. With the exception of OEF, values differed between 2D PET and 3D PET. However, average gray matter values of scatter-corrected 3D PET were comparable to those of 2D PET: 55 +/- 11 (mL/100 mL/min), 3.7 +/- 0.5 (mL/100 mL), 3.8 +/- 0.7 (mL/100 mL/min), and 0.36 +/- 0.06, respectively. Even though the 2 PET scanners with different crystal materials, data acquisition systems, spatial resolution, and attenuation-correction methods were used, the agreement of the results between 2D PET and scatter-corrected 3D PET was excellent.
Conclusion: Scatter coincidence is a problem in 3D PET for quantitative (15)O study. The combination of both the present PET/CT device and the HDE scatter correction permits quantitative 3D PET with the same degree of accuracy as 2D PET and with a lower radiation dose. The present scanner is also applicable to conventional steady-state (15)O gas inhalation if inhaled doses are adjusted appropriately.
Similar articles
-
Quantitative evaluation of cerebral blood flow and oxygen metabolism in normal anesthetized rats: 15O-labeled gas inhalation PET with MRI Fusion.J Nucl Med. 2013 Feb;54(2):283-90. doi: 10.2967/jnumed.112.109751. Epub 2013 Jan 3. J Nucl Med. 2013. PMID: 23287575
-
Evaluation of dynamic row-action maximum likelihood algorithm reconstruction for quantitative 15O brain PET.Ann Nucl Med. 2009 Sep;23(7):627-38. doi: 10.1007/s12149-009-0280-2. Epub 2009 Jun 30. Ann Nucl Med. 2009. PMID: 19562437
-
Validation of a simplified scatter correction method for 3D brain PET with 15O.Ann Nucl Med. 2016 Dec;30(10):690-698. doi: 10.1007/s12149-016-1114-7. Epub 2016 Aug 17. Ann Nucl Med. 2016. PMID: 27534771 Free PMC article.
-
Human cerebral circulation: positron emission tomography studies.Ann Nucl Med. 2005 Apr;19(2):65-74. doi: 10.1007/BF03027383. Ann Nucl Med. 2005. PMID: 15909484 Review.
-
In-vivo positron emission tomography (PET) measurement of cerebral oxygen metabolism in small animals.Yakugaku Zasshi. 2008 Sep;128(9):1267-73. doi: 10.1248/yakushi.128.1267. Yakugaku Zasshi. 2008. PMID: 18758140 Review.
Cited by
-
A generic bioheat transfer thermal model for a perfused tissue.J Biomech Eng. 2009 Jul;131(7):074506. doi: 10.1115/1.3127260. J Biomech Eng. 2009. PMID: 19640142 Free PMC article.
-
Quantitative hemodynamic PET imaging using image-derived arterial input function and a PET/MR hybrid scanner.J Cereb Blood Flow Metab. 2017 Apr;37(4):1435-1446. doi: 10.1177/0271678X16656200. Epub 2016 Jan 1. J Cereb Blood Flow Metab. 2017. PMID: 27401805 Free PMC article.
-
Oxygen-15 labeled CO2, O2, and CO PET in small animals: evaluation using a 3D-mode microPET scanner and impact of reconstruction algorithms.EJNMMI Res. 2017 Oct 27;7(1):91. doi: 10.1186/s13550-017-0335-7. EJNMMI Res. 2017. PMID: 29080056 Free PMC article.
-
Imaging of Moyamoya Disease and Moyamoya Syndrome: Current Status.J Comput Assist Tomogr. 2019 Mar/Apr;43(2):257-263. doi: 10.1097/RCT.0000000000000834. J Comput Assist Tomogr. 2019. PMID: 30589721 Free PMC article. Review.
-
In vivo positron emission tomographic blood pool imaging in an immunodeficient mouse model using 18F-fluorodeoxyglucose labeled human erythrocytes.PLoS One. 2019 Jan 25;14(1):e0211012. doi: 10.1371/journal.pone.0211012. eCollection 2019. PLoS One. 2019. PMID: 30682160 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources