An investigation of the validity of the Lifeshirt in comparison to standard polysomnography in the detection of obstructive sleep apnea
- PMID: 18083629
- DOI: 10.1016/j.sleep.2007.11.003
An investigation of the validity of the Lifeshirt in comparison to standard polysomnography in the detection of obstructive sleep apnea
Abstract
Background: Due to the cost and waiting times for traditional sleep studies, there is great interest in finding alternatives for the diagnosis of sleep apnea. Several alternatives exist, including ambulatory devices. Our laboratory utilized one such device, a form-fitting vest called the Lifeshirt, and compared its accuracy to that of polysomnography (PSG).
Methods: PSG was performed with simultaneous utilization of the Lifeshirt on 50 individuals who met screening criteria for obstructive sleep apnea. Participants came to the sleep laboratory approximately 2 h before their normal bedtime. A sleep technician prepared each participant for PSG and Lifeshirt monitoring. "Lights Out" occurred when the participant was ready for bed, and time in bed was standardized to approximately 7 h. PSGs were scored by experienced personnel in our laboratory, while Lifeshirt data were sent electronically to the Lifeshirt manufacturer for analysis. The major variable that was compared between PSG and Lifeshirt was the apnea hypopnea index (AHI), or the number of apneas and hypopneas per hour. Due to incomplete data on two participants, analyses were completed on 48 individuals.
Results: Sensitivity of the Lifeshirt ranged from .85 (AHI of > or =5) to 1.00 (AHI of > or =25). Specificity ranged from .67 to 1.00. Using the Bland-Altman technique of determining agreement, the mean difference between the AHI of the Lifeshirt and PSG was 1.02 (+/-16.36). When these values are plotted, every case falls within the limits of agreement, with one exception.
Conclusions: The Lifeshirt compared favorably with PSG and could be used with considerable confidence for the screening of patients with suspected obstructive sleep apnea.
Similar articles
-
[Value of oxygen saturation combined airflow in diagnosis of obstructive sleep apnea hypopnea syndrome].Zhonghua Yi Xue Za Zhi. 2008 Sep 9;88(34):2395-8. Zhonghua Yi Xue Za Zhi. 2008. PMID: 19087713 Chinese.
-
[Evaluation of a portable device based on peripheral arterial tone in the detection of obstructive sleep apnea].Zhonghua Er Bi Yan Hou Tou Jing Wai Ke Za Zhi. 2012 Feb;47(2):112-6. Zhonghua Er Bi Yan Hou Tou Jing Wai Ke Za Zhi. 2012. PMID: 22455808 Chinese.
-
Obstructive sleep apnoea syndrome: comparison between polysomnography and portable sleep monitoring based on jaw recordings.Neurophysiol Clin. 2011 Oct;41(4):191-8. doi: 10.1016/j.neucli.2011.09.001. Epub 2011 Oct 5. Neurophysiol Clin. 2011. PMID: 22078731
-
Portable monitoring in the diagnosis of the obstructive sleep apnea syndrome.Semin Respir Crit Care Med. 2005 Feb;26(1):56-67. doi: 10.1055/s-2005-864200. Semin Respir Crit Care Med. 2005. PMID: 16052418 Review.
-
Contemporary issues in the diagnosis and treatment of obstructive sleep apnea.Postgrad Med. 2008 Jul;120(2):46-52. doi: 10.3810/pgm.2008.07.1790. Postgrad Med. 2008. PMID: 18654068 Review.
Cited by
-
Design and evaluation of a ubiquitous chest-worn cardiopulmonary monitoring system for healthcare application: a pilot study.Med Biol Eng Comput. 2017 Feb;55(2):283-294. doi: 10.1007/s11517-016-1518-5. Epub 2016 May 13. Med Biol Eng Comput. 2017. PMID: 27177545
-
Laboratory Validation of Hexoskin Biometric Shirt at Rest, Submaximal Exercise, and Maximal Exercise While Riding a Stationary Bicycle.J Occup Environ Med. 2019 Apr;61(4):e104-e111. doi: 10.1097/JOM.0000000000001537. J Occup Environ Med. 2019. PMID: 30946696 Free PMC article.
-
Sleep apnea monitoring and diagnosis based on pulse oximetry and tracheal sound signals.Med Biol Eng Comput. 2010 Nov;48(11):1087-97. doi: 10.1007/s11517-010-0674-2. Epub 2010 Aug 24. Med Biol Eng Comput. 2010. PMID: 20734154
-
Utility of a smartphone based system (cvrphone) to accurately determine apneic events from electrocardiographic signals.PLoS One. 2019 Jun 17;14(6):e0217217. doi: 10.1371/journal.pone.0217217. eCollection 2019. PLoS One. 2019. PMID: 31206522 Free PMC article.
-
Theory and developments in an unobtrusive cardiovascular system representation: ballistocardiography.Open Biomed Eng J. 2010;4:201-16. doi: 10.2174/1874120701004010201. Epub 2010 Oct 10. Open Biomed Eng J. 2010. PMID: 21673836 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous