Remelting of highly cross-linked polyethylene worn under laboratory conditions
- PMID: 18090470
- DOI: 10.1097/BLO.0b013e318157601c
Remelting of highly cross-linked polyethylene worn under laboratory conditions
Abstract
It has been suggested that apparent wear damage in highly cross-linked polyethylene acetabular liners can be removed with subsequent remelting of retrieved liners. To test this hypothesis, we remelted liners that had been previously tested under controlled laboratory conditions and that had experienced nonzero wear rates and visible wear damage. Five liner groups were examined: three with a range of irradiation doses without heat treatment, and two irradiated to 100 kGys, one with remelting, and the other annealing. All groups had been worn in a hip simulator under impingement conditions that produced nonzero wear rates and loss of machining marks. Each liner was cut into quadrants that were graded for wear damage before and after posttest remelting. Cross-linked liners not previously heat-treated lost all prior damage and all machining marks. Remelted liners and three of six annealed liners experienced only a slight return of machining marks at the interface of the burnished area and the remaining intact machining marks. Our experiments represent a severe wear case but demonstrate removal of material from the surface through measurable wear prevents the return of identifiable machining marks despite remelting.
Similar articles
-
Polyethylene cross-linking by two different methods reduces acetabular liner wear in a hip joint wear simulator.J Orthop Res. 2003 Sep;21(5):761-6. doi: 10.1016/S0736-0266(02)00240-1. J Orthop Res. 2003. PMID: 12919860
-
Surface-gradient cross-linked polyethylene acetabular cups: oxidation resistance and wear against smooth and rough femoral balls.Clin Orthop Relat Res. 2005 Jan;(430):80-8. Clin Orthop Relat Res. 2005. PMID: 15662307
-
Wear behaviour of cross-linked polyethylene assessed in vitro under severe conditions.Biomaterials. 2005 Jun;26(16):3259-67. doi: 10.1016/j.biomaterials.2004.07.070. Biomaterials. 2005. PMID: 15603821
-
Conventional ultra-high molecular weight polyethylene: a gold standard of sorts.Instr Course Lect. 2005;54:183-7. Instr Course Lect. 2005. PMID: 15948445 Review.
-
Highly cross-linked polyethylene bearing surfaces in total hip arthroplasty.J Surg Orthop Adv. 2008 Spring;17(1):27-33. J Surg Orthop Adv. 2008. PMID: 18284901 Review.
Cited by
-
Effects of Anticipation and Dual-Tasking on Lower Limb Biomechanics While Performing Change-of-Direction Tasks in Physically Active Individuals: A Systematic Review with Meta-Analysis.Sports Med. 2025 Apr;55(4):857-876. doi: 10.1007/s40279-025-02182-w. Epub 2025 Mar 20. Sports Med. 2025. PMID: 40111742 Free PMC article.
-
Retrieved highly crosslinked UHMWPE acetabular liners have similar wear damage as conventional UHMWPE.Clin Orthop Relat Res. 2011 Feb;469(2):387-94. doi: 10.1007/s11999-010-1552-9. Clin Orthop Relat Res. 2011. PMID: 20844998 Free PMC article.
-
Scratching vulnerability of conventional vs highly cross-linked polyethylene liners because of large embedded third-body particles.J Arthroplasty. 2012 May;27(5):742-9. doi: 10.1016/j.arth.2011.10.002. Epub 2011 Nov 23. J Arthroplasty. 2012. PMID: 22115764 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources