Interaural time and level difference thresholds for acoustically presented signals in post-lingually deafened adults fitted with bilateral cochlear implants using CIS+ processing
- PMID: 18091105
- DOI: 10.1097/AUD.0b013e31815d636f
Interaural time and level difference thresholds for acoustically presented signals in post-lingually deafened adults fitted with bilateral cochlear implants using CIS+ processing
Abstract
Objectives: The main purpose of the study was to measure thresholds for interaural time differences (ITDs) and interaural level differences (ILDs) for acoustically presented noise signals in adults with bilateral cochlear implants (CIs). A secondary purpose was to assess the correlation between the ILD and ITD thresholds and error scores in a horizontal-plane localization task, to test the hypothesis that localization by individuals with bilateral implants is mediated by the processing of ILD cues.
Design: Eleven adults, all postlingually deafened and all bilaterally fitted with MED-EL COMBI 40+ CIs, were tested in ITD and ILD discrimination tasks in which signals were presented acoustically through headphones that fit over their two devices. The stimulus was a 200-msec burst of Gaussian noise bandpass filtered from 100 to 4000 Hz. A two-interval forced-choice adaptive procedure was used in which the subject had to respond on each trial whether the lateral positions of the two sound images (with the interaural difference favoring the left and right sides in the two intervals) moved from left-to-right or right-to-left.
Results: In agreement with previously reported data, ITD thresholds for the subjects with bilateral implants were poor. The best threshold was approximately 400 microsec, and only five of 11 subjects tested achieved thresholds <1000 microsec. In contrast, ILD thresholds were relatively good; mean threshold was 3.8 dB with the initial compression circuit on the implant devices activated and 1.9 dB with the compression deactivated. The ILD and ITD thresholds were higher than previously reported thresholds obtained with direct electrical stimulation (generally, <1.0 dB and 100 to 200 microsec, respectively). When the data from two outlying subjects were omitted, ILD thresholds were highly correlated with total error score in a horizontal-plane localization task, computed for sources near midline (r = 0.87, p < 0.01).
Conclusions: The higher ILD and ITD thresholds obtained in this study with acoustically presented signals (when compared with prior data with direct electrical stimulation) can be attributed-at least partially-to the signal processing carried out by the CI in the former case. The processing strategy effectively leaves only envelope information as a basis for ITD discrimination, which, for the acoustically presented noise stimuli, is mainly coded in the onset information. The operation of the compression circuit reduces the ILDs in the signal, leading to elevated ILD thresholds for the acoustically presented signals in this condition. The large magnitude of the ITD thresholds indicates that ITDs could not have contributed to the performance in the horizontal-plane localization task. Overall, the results suggest that for subjects using bilateral implants, localization of noise signals is mediated entirely by ILD cues, with little or no contribution from ITD information.
Similar articles
-
Horizontal-plane localization of noise and speech signals by postlingually deafened adults fitted with bilateral cochlear implants.Ear Hear. 2007 Aug;28(4):524-41. doi: 10.1097/AUD.0b013e31806dc21a. Ear Hear. 2007. PMID: 17609614
-
Lateralization of interimplant timing and level differences in children who use bilateral cochlear implants.Ear Hear. 2010 Aug;31(4):441-56. doi: 10.1097/AUD.0b013e3181d4f228. Ear Hear. 2010. PMID: 20489647
-
Effects of interaural pitch matching and auditory image centering on binaural sensitivity in cochlear implant users.Ear Hear. 2015 May-Jun;36(3):e62-8. doi: 10.1097/AUD.0000000000000135. Ear Hear. 2015. PMID: 25565660 Free PMC article.
-
Perception and coding of interaural time differences with bilateral cochlear implants.Hear Res. 2015 Apr;322:138-50. doi: 10.1016/j.heares.2014.10.004. Epub 2014 Oct 19. Hear Res. 2015. PMID: 25456088 Review.
-
Auditory Model-Based Sound Direction Estimation With Bilateral Cochlear Implants.Trends Hear. 2015 Dec 1;19:2331216515616378. doi: 10.1177/2331216515616378. Trends Hear. 2015. PMID: 26631106 Free PMC article. Review.
Cited by
-
Cochlear implant patients' localization using interaural level differences exceeds that of untrained normal hearing listeners.J Acoust Soc Am. 2012 May;131(5):EL382-7. doi: 10.1121/1.3699017. J Acoust Soc Am. 2012. PMID: 22559456 Free PMC article.
-
Temporal weighting functions for interaural time and level differences. II. The effect of binaurally synchronous temporal jitter.J Acoust Soc Am. 2011 Jan;129(1):293-300. doi: 10.1121/1.3514422. J Acoust Soc Am. 2011. PMID: 21303010 Free PMC article.
-
Microsecond interaural time difference discrimination restored by cochlear implants after neonatal deafness.Elife. 2021 Jan 11;10:e59300. doi: 10.7554/eLife.59300. Elife. 2021. PMID: 33427644 Free PMC article.
-
The perception of the stereo effect in bilateral and bimodal cochlear implant users and its contribution to music enjoyment.PLoS One. 2020 Jul 6;15(7):e0235435. doi: 10.1371/journal.pone.0235435. eCollection 2020. PLoS One. 2020. PMID: 32628690 Free PMC article.
-
Interaural time difference sensitivity under binaural cochlear implant stimulation persists at high pulse rates up to 900 pps.Sci Rep. 2023 Mar 7;13(1):3785. doi: 10.1038/s41598-023-30569-0. Sci Rep. 2023. PMID: 36882473 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous