Spectral optical coherence tomography in video-rate and 3D imaging of contact lens wear
- PMID: 18091301
- DOI: 10.1097/OPX.0b013e31815b9e0e
Spectral optical coherence tomography in video-rate and 3D imaging of contact lens wear
Abstract
Purpose: To present the applicability of spectral optical coherence tomography (SOCT) for video-rate and three-dimensional imaging of a contact lens on the eye surface.
Methods: The SOCT prototype instrument constructed at Nicolaus Copernicus University (Torun, Poland) is based on Fourier domain detection, which enables high sensitivity (96 dB) and increases the speed of imaging 60 times compared with conventional optical coherence tomography techniques. Consequently, video-rate imaging and three-dimensional reconstructions can be achieved, preserving the high quality of the image. The instrument operates under clinical conditions in the Ophthalmology Department (Collegium Medicum Nicolaus Copernicus University, Bydgoszcz, Poland). A total of three eyes fitted with different contact lenses were examined with the aid of the instrument. Before SOCT measurements, slit lamp examinations were performed.
Results: Data, which are representative for each imaging mode, are presented. The instrument provided high-resolution (4 microm axial x 10 microm transverse) tomograms with an acquisition time of 40 micros per A-scan. Video-rate imaging allowed the simultaneous quantitative evaluation of the movement of the contact lens and assessment of the fitting relationship between the lens and the ocular surface. Three-dimensional scanning protocols further improved lens visualization and fit evaluation.
Conclusions: SOCT allows video-rate and three-dimensional cross-sectional imaging of the eye fitted with a contact lens. The analysis of both imaging modes suggests the future applicability of this technology to the contact lens field.
Similar articles
-
Spectral optical coherence tomography: a new imaging technique in contact lens practice.Ophthalmic Physiol Opt. 2006 Mar;26(2):127-32. doi: 10.1111/j.1475-1313.2006.00371.x. Ophthalmic Physiol Opt. 2006. PMID: 16460312
-
Imaging of the lens capsule with an ultrahigh-resolution spectral optical coherence tomography prototype based on a femtosecond laser.Br J Ophthalmol. 2010 Mar;94(3):275-7. doi: 10.1136/bjo.2008.155879. Br J Ophthalmol. 2010. PMID: 20215371
-
Spectral optical coherence tomography: a novel technique for cornea imaging.Cornea. 2006 Sep;25(8):960-5. doi: 10.1097/01.ico.0000224644.81719.59. Cornea. 2006. PMID: 17102675
-
[Optical coherence tomography: from retina imaging to intraoperative use - a review].Klin Monbl Augenheilkd. 2009 Dec;226(12):958-64. doi: 10.1055/s-0028-1109939. Epub 2009 Dec 15. Klin Monbl Augenheilkd. 2009. PMID: 20108189 Review. German.
-
[Application of spectral optical coherent tomography (SOCT) in ophthalmology].Klin Oczna. 2004;106(4-5):699-702. Klin Oczna. 2004. PMID: 15646498 Review. Polish.
Cited by
-
Applications of Anterior Segment Optical Coherence Tomography in Cornea and Ocular Surface Diseases.J Ophthalmol. 2016;2016:4971572. doi: 10.1155/2016/4971572. Epub 2016 Sep 19. J Ophthalmol. 2016. PMID: 27721988 Free PMC article. Review.
-
Full-range imaging of eye accommodation by high-speed long-depth range optical frequency domain imaging.Biomed Opt Express. 2010 Nov 23;1(5):1491-1501. doi: 10.1364/BOE.1.001491. Biomed Opt Express. 2010. PMID: 21258564 Free PMC article.
-
Spectral optical coherence tomography vs. fluorescein pattern for rigid gas-permeable lens fit.Med Sci Monit. 2014 Jul 4;20:1137-41. doi: 10.12659/MSM.890269. Med Sci Monit. 2014. PMID: 24995686 Free PMC article.
-
In situ visualization of tears on contact lens using ultra high resolution optical coherence tomography.Eye Contact Lens. 2009 Mar;35(2):44-9. doi: 10.1097/ICL.0b013e31819579f2. Eye Contact Lens. 2009. PMID: 19265323 Free PMC article.
-
Entire contact lens imaged in vivo and in vitro with spectral domain optical coherence tomography.Eye Contact Lens. 2010 Mar;36(2):73-6. doi: 10.1097/ICL.0b013e3181d0423e. Eye Contact Lens. 2010. PMID: 20093938 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous