Jumbo forceps are superior to standard large-capacity forceps in obtaining diagnostically adequate inflammatory bowel disease surveillance biopsy specimens

Affiliations
Comparative Study

Jumbo forceps are superior to standard large-capacity forceps in obtaining diagnostically adequate inflammatory bowel disease surveillance biopsy specimens

B Joseph Elmunzer et al. Gastrointest Endosc. 2008 Aug.

Abstract

Background: In inflammatory bowel disease (IBD) surveillance colonoscopy, an increased number of biopsy specimens correlates with a higher dysplasia detection rate. Larger biopsy specimens may also increase the diagnostic yield.

Objective: To compare a new jumbo forceps with a standard large-capacity forceps in obtaining diagnostically adequate IBD surveillance biopsy specimens.

Design: Prospective single-center study.

Patients and methods: Twenty-four patients who were undergoing an IBD surveillance colonoscopy were enrolled. As part of standard IBD surveillance, 8 paired biopsy specimens were obtained from the rectosigmoid by using the jumbo forceps and a standard large-capacity forceps.

Outcome measurements: Biopsy specimens were deemed adequate if they met all 3 of the following criteria: (1) length > or =3 mm, (2) penetration into the muscularis mucosa, and (3) < 20% crush artifact.

Results: The proportion of adequate biopsy specimens obtained with the jumbo forceps was significantly higher than that obtained with the large-capacity control forceps (67% vs 48%, P < .0001). The average length of the biopsy specimen obtained with the jumbo forceps was 4.00 mm (95% CI, 3.81-4.20 mm) compared with 3.19 mm (95% CI, 2.99-3.38 mm) with the large-capacity (control) forceps.

Limitations: (1) No validated outcome measurement for the quality of GI biopsy specimens exists and (2) in this study, interobserver variability between pathologists was high.

Conclusions: The jumbo forceps was superior to a standard large-capacity forceps in obtaining diagnostically adequate IBD surveillance biopsy specimens. Because biopsy specimens obtained with the jumbo forceps were larger, the use of this forceps for IBD surveillance will allow the endoscopist to sample a larger colonic mucosal surface area, potentially resulting in an increased dysplasia detection rate.