Improved accuracy of component alignment with the implementation of image-free navigation in total knee arthroplasty
- PMID: 18157493
- DOI: 10.1007/s00167-007-0420-y
Improved accuracy of component alignment with the implementation of image-free navigation in total knee arthroplasty
Abstract
Accuracy of implant positioning and reconstruction of the mechanical leg axis are major requirements for achieving good long-term results in total knee arthroplasty (TKA). The purpose of the present study was to determine whether image-free computer navigation technology has the potential to improve the accuracy of component alignment in TKA cohorts of experienced surgeons immediately and constantly. One hundred patients with primary arthritis of the knee underwent the unilateral total knee arthroplasty. The cohort of 50 TKAs implanted with conventional instrumentation was directly followed by the cohort of the very first 50 computer-assisted TKAs. All surgeries were performed by two senior surgeons. All patients received the Zimmer NexGen total knee prosthesis (Zimmer Inc., Warsaw, IN, USA). There was no variability regarding surgeons or surgical technique, except for the use of the navigation system (StealthStation) Treon plus Medtronic Inc., Minnesota, MI, USA). Accuracy of implant positioning was measured on postoperative long-leg standing radiographs and standard lateral X-rays with regard to the valgus angle and the coronal and sagittal component angles. In addition, preoperative deformities of the mechanical leg axis, tourniquet time, age, and gender were correlated. Statistical analyses were performed using the SPSS 15.0 (SPSS Inc., Chicago, IL, USA) software package. Independent t-tests were used, with significance set at P < 0.05 (two-tailed) to compare differences in mean angular values and frontal mechanical alignment between the two cohorts. To compute the rate of optimally implanted prostheses between the two groups we used the chi(2) test. The average postoperative radiological frontal mechanical alignment was 1.88 degrees of varus (range 6.1 degrees of valgus-10.1 degrees of varus; SD 3.68 degrees ) in the conventional cohort and 0.28 degrees of varus (range 3.7 degrees -6.0 degrees of varus; SD 1.97 degrees ) in the navigated cohort. Including all criteria for optimal implant alignment, 16 cases (32%) in the conventional cohort and 31 cases (62%) in the navigated cohort have been implanted optimally. The average difference in tourniquet time was modest with additional 12.9 min in the navigated cohort compared to the conventional cohort. Our findings suggest that the experienced knee surgeons can improve immediately and constantly the accuracy of component orientation using an image-free computer-assisted navigation system in TKA. The computer-assisted technology has shown to be easy to use, safe, and efficient in routine knee replacement surgery. We believe that navigation is a key technology for various current and future surgical alignment topics and minimal-invasive lower limb surgery.
Similar articles
-
The immediate effect of navigation on implant accuracy in primary mini-invasive unicompartmental knee arthroplasty.Knee Surg Sports Traumatol Arthrosc. 2008 Dec;16(12):1133-40. doi: 10.1007/s00167-008-0618-7. Epub 2008 Sep 20. Knee Surg Sports Traumatol Arthrosc. 2008. PMID: 18810391
-
Accuracy of Computed Tomography-Based Navigation-Assisted Total Knee Arthroplasty: Outlier Analysis.J Arthroplasty. 2017 Jan;32(1):47-52. doi: 10.1016/j.arth.2016.05.069. Epub 2016 Jun 7. J Arthroplasty. 2017. PMID: 27369304
-
[Computer-assisted surgery in total knee replacement. Preliminary results: report of 60 cases].Rev Chir Orthop Reparatrice Appar Mot. 2008 May;94(3):261-7. doi: 10.1016/j.rco.2007.10.005. Epub 2008 Feb 21. Rev Chir Orthop Reparatrice Appar Mot. 2008. PMID: 18456061 French.
-
Imageless navigation system does not improve component rotational alignment in total knee arthroplasty.J Surg Res. 2011 Dec;171(2):590-600. doi: 10.1016/j.jss.2010.05.006. Epub 2010 Oct 31. J Surg Res. 2011. PMID: 21176919 Review.
-
Current Role of Computer Navigation in Total Knee Arthroplasty.J Arthroplasty. 2018 Jul;33(7):1989-1993. doi: 10.1016/j.arth.2018.01.027. Epub 2018 Jan 31. J Arthroplasty. 2018. PMID: 29506932 Review.
Cited by
-
Efficacy and reliability of active robotic-assisted total knee arthroplasty compared with conventional total knee arthroplasty: a systematic review and meta-analysis.Postgrad Med J. 2019 Mar;95(1121):125-133. doi: 10.1136/postgradmedj-2018-136190. Epub 2019 Feb 26. Postgrad Med J. 2019. PMID: 30808721 Free PMC article.
-
Computer-assisted surgery can reduce blood loss after total knee arthroplasty.Knee Surg Sports Traumatol Arthrosc. 2009 Apr;17(4):356-60. doi: 10.1007/s00167-008-0683-y. Epub 2008 Dec 16. Knee Surg Sports Traumatol Arthrosc. 2009. PMID: 19083205 Clinical Trial.
-
One year after navigated total knee replacement, no clinically relevant difference found between fixed bearing and mobile bearing knee replacement in a double-blind randomized controlled trial.Open Orthop J. 2011;5:201-8. doi: 10.2174/1874325001105010201. Epub 2011 May 27. Open Orthop J. 2011. PMID: 21687563 Free PMC article.
-
Augmented reality-based navigation system applied to tibial bone resection in total knee arthroplasty.J Exp Orthop. 2019 Nov 11;6(1):44. doi: 10.1186/s40634-019-0212-6. J Exp Orthop. 2019. PMID: 31712907 Free PMC article.
-
A comparison of utilization and short-term complications of technology-assisted versus conventional total knee arthroplasty.Knee Surg Relat Res. 2022 Mar 18;34(1):14. doi: 10.1186/s43019-022-00143-5. Knee Surg Relat Res. 2022. PMID: 35303957 Free PMC article.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical