Comparison of central corneal thickness as measured by non-contact specular microscopy and ultrasound pachymetry before and post LASIK
- PMID: 18173409
- DOI: 10.1111/j.1442-9071.2007.01633.x
Comparison of central corneal thickness as measured by non-contact specular microscopy and ultrasound pachymetry before and post LASIK
Abstract
Purpose: To compare central corneal thickness (CCT) measurements made by ultrasound pachymetry and non-contact specular microscopy in eyes before and post laser in situ keratomileusis (LASIK).
Methods: Forty eyes of 20 consecutive refractive surgery candidates were included in this study. The CCT in both eyes was measured before and 3 months after LASIK with a non-contact specular microscope and a ultrasound pachymeter.
Result: Both before and after LASIK, the difference of the mean CCT between non-contact specular microscopy and ultrasound pachymetry was statistically significant (P < 0.001 before surgery, and P = 0.02 after surgery). The difference between them was -14.41 microm and -3.69 microm, respectively. In both situations, non-contact specular microscopy and ultrasound pachymetry measurements were highly correlated (r = 0.959, before surgery, and r = 0.979 after LASIK). Bland-Altman analysis showed that either before or post LASIK, the 95% limits of agreement have a span of approximately 30 microm. Preoperatively, the mean standard deviation of repeated measurements was 4.26 microm by ultrasound pachymetry, and 4.02 microm by non-contact specular microscopy, whereas it was 3.83 microm by ultrasound pachymetry, and 3.89 microm by non-contact specular microscopy, postoperatively. No statistically significant differences were found between the mean standard deviations by these two devices in both situations (P = 0.561 before surgery, and P = 0.849 after LASIK).
Conclusions: Both before and after LASIK, non-contact specular microscopy is not in complete agreement with ultrasound pachymetry in measuring CCT. Both devices provided comparable repeatability of measurements.
Similar articles
-
Comparison of corneal thickness measurements using Orbscan II, non-contact specular microscopy, and ultrasonic pachymetry in eyes after laser in situ keratomileusis.Br J Ophthalmol. 2004 Apr;88(4):466-8. doi: 10.1136/bjo.2003.030361. Br J Ophthalmol. 2004. PMID: 15031156 Free PMC article.
-
Central and peripheral corneal thickness measured with optical coherence tomography, Scheimpflug imaging, and ultrasound pachymetry in normal, keratoconus-suspect, and post-laser in situ keratomileusis eyes.J Cataract Refract Surg. 2009 Jun;35(6):1055-62. doi: 10.1016/j.jcrs.2009.01.022. J Cataract Refract Surg. 2009. PMID: 19465292
-
Comparison of corneal thickness measurements using ultrasound and Orbscan slit-scanning topography in normal and post-LASIK eyes.J Cataract Refract Surg. 2001 Nov;27(11):1823-8. doi: 10.1016/s0886-3350(01)01089-6. J Cataract Refract Surg. 2001. PMID: 11709257
-
Corneal pachymetry mapping with high-speed optical coherence tomography.Ophthalmology. 2006 May;113(5):792-9.e2. doi: 10.1016/j.ophtha.2006.01.048. Ophthalmology. 2006. PMID: 16650675 Free PMC article.
-
Non-contact specular microscopy with Topcon instruments to assess central corneal thickness of healthy human eyes - A 20 year review.Cont Lens Anterior Eye. 2021 Aug;44(4):101385. doi: 10.1016/j.clae.2020.11.005. Epub 2020 Dec 16. Cont Lens Anterior Eye. 2021. PMID: 33341365 Review.
Cited by
-
Corneal Thickness Response after Anesthetic Eye Drops: Our Own Results and Meta-Analysis.Biomed Res Int. 2018 Mar 5;2018:4743721. doi: 10.1155/2018/4743721. eCollection 2018. Biomed Res Int. 2018. PMID: 29693008 Free PMC article.
-
Age-related changes in central corneal thickness in normal eyes among the adult Lithuanian population.Clin Interv Aging. 2014 Jul 16;9:1145-51. doi: 10.2147/CIA.S61790. eCollection 2014. Clin Interv Aging. 2014. PMID: 25075183 Free PMC article.
-
Effects of inhaled fluticasone on intraocular pressure and central corneal thickness in asthmatic children without a family history of glaucoma.Middle East Afr J Ophthalmol. 2012 Jul-Sep;19(3):314-9. doi: 10.4103/0974-9233.97936. Middle East Afr J Ophthalmol. 2012. PMID: 22837626 Free PMC article.
-
Comparison of central corneal thickness measurements by Pentacam, noncontact specular microscope, and ultrasound pachymetry in normal and post-LASIK eyes.Saudi J Ophthalmol. 2009 Oct;23(3-4):181-7. doi: 10.1016/j.sjopt.2009.10.002. Epub 2009 Oct 24. Saudi J Ophthalmol. 2009. PMID: 23960858 Free PMC article.
-
Comparison of specular microscopy and ultrasound pachymetry before and after cataract surgery.Graefes Arch Clin Exp Ophthalmol. 2017 Feb;255(2):387-392. doi: 10.1007/s00417-016-3537-1. Epub 2016 Nov 22. Graefes Arch Clin Exp Ophthalmol. 2017. PMID: 27878429
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources