Evaluation of an acid-fast bacilli autostainer for concentrated sputum smears
- PMID: 18173875
Evaluation of an acid-fast bacilli autostainer for concentrated sputum smears
Abstract
Setting: An automatic acid-fast bacilli (AFB) stainer was developed to improve laboratory workflow and protect laboratory workers.
Objective: To determine the usefulness of the 7720 Aerospray AFB slide stainer for liquefied and concentrated respiratory specimens.
Design: With 24 smear-positive specimens submitted for AFB stain and culture in a tertiary care hospital, the autostainer was evaluated to determine whether smears would be sloughed from slides during the staining process by loading distilled water instead of staining reagents. With 109 more specimens, the performance of the 7720 Aerospray was compared with manual florescence staining as a gold standard. The tested smears were prepared with Mycohold treatment.
Results: In most cases, the slides exposed to the autostainer containing distilled water showed lower grades of AFB than those slides not exposed to the autostainer (14/24 in Mycohold-treated slides). Almost half (54/109, 49.5%) of the Mycohold-treated autostained smears had a lower grade than the same smears stained manually. Among 62 Mycohold-treated autostained smear-negative specimens, respectively eight (12.9%) and 19 (30.6%) were high and low false-negatives.
Conclusion: The 7720 Aerospray showed poor performance in AFB staining of concentrated sputum specimens.
Similar articles
-
Bulk acid-fast staining of sputum smears: time to end a taboo.Int J Tuberc Lung Dis. 2009 Sep;13(9):1119-23. Int J Tuberc Lung Dis. 2009. PMID: 19723401
-
Rapid detection of smear-negative Mycobacterium tuberculosis by PCR and sequencing for rifampin resistance with DNA extracted directly from slides.J Clin Microbiol. 2001 Jan;39(1):51-2. doi: 10.1128/JCM.39.1.51-52.2001. J Clin Microbiol. 2001. PMID: 11136747 Free PMC article.
-
Reducing the number of sputum samples examined and thresholds for positivity: an opportunity to optimise smear microscopy.Int J Tuberc Lung Dis. 2007 Sep;11(9):953-8. Int J Tuberc Lung Dis. 2007. PMID: 17705971
-
Factors affecting the clinical value of microscopy for acid-fast bacilli.Rev Infect Dis. 1984 Mar-Apr;6(2):214-22. doi: 10.1093/clinids/6.2.214. Rev Infect Dis. 1984. PMID: 6203161 Review.
-
Critical appraisal of current recommendations and practices for tuberculosis sputum smear microscopy.Int J Tuberc Lung Dis. 2007 Sep;11(9):946-52. Int J Tuberc Lung Dis. 2007. PMID: 17705970
Cited by
-
Performance of Bleach Method Sputum Smear Microscopy for the Diagnosis of Tuberculosis in a Highly Endemic District in Lima, Peru.Int J Environ Res Public Health. 2022 Dec 22;20(1):135. doi: 10.3390/ijerph20010135. Int J Environ Res Public Health. 2022. PMID: 36612454 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources