Assessment of stroke volume index with three different bioimpedance algorithms: lack of agreement compared to thermodilution
- PMID: 18188539
- PMCID: PMC2271085
- DOI: 10.1007/s00134-007-0938-y
Assessment of stroke volume index with three different bioimpedance algorithms: lack of agreement compared to thermodilution
Abstract
Objective: The accuracy of bioimpedance stroke volume index (SVI) is questionable as studies report inconsistent results. It remains unclear whether the algorithms alone are responsible for these findings. We analyzed the raw impedance data with three algorithms and compared bioimpedance SVI to transpulmonary thermodilution (SVI(TD)).
Design and setting: Prospective observational clinical study in a university hospital.
Patients: Twenty adult patients scheduled for coronary artery bypass grafting (CABG).
Interventions: SVI(TD) and bioimpedance parameters were simultaneously obtained before surgery (t1), after bypass (t2), after sternal closure (t3), at the intensive care unit (t4), at normothermia (t5), after extubation (t6) and before discharge (t7). Bioimpedance data were analyzed off-line using cylinder (Kubicek: SVI(K); Wang: SVI(W)) and truncated cone based algorithms (Sramek-Bernstein: SVI(SB)).
Measurements and results: Bias and precision between the SVI(TD) and SVI(K), SVI(SB), and SVI(W) was 1.0+/-10.8, 9.8+/-11.4, and -15.7+/-8.2 ml/m2 respectively, while the mean error was abundantly above 30%. Analysis of data per time moment resulted in a mean error above 30%, except for SVI(W) at t2 (28%).
Conclusions: Estimation of SVI by cylinder or truncated cone based algorithms is not reliable for clinical decision making in patients undergoing CABG surgery. A more robust approach for estimating bioimpedance based SVI may exclude inconsistencies in the underlying algorithms in existing thoracic bioimpedance cardiography devices.
Figures
References
-
- {'text': '', 'ref_index': 1, 'ids': [{'type': 'PubMed', 'value': '5434111', 'is_inner': True, 'url': 'https://pubmed.ncbi.nlm.nih.gov/5434111/'}]}
- Swan HJ, Ganz W, Forrester J, Marcus H, Diamond G, Chonette D (1970) Catheterization of the heart in man with use of a flow-directed balloon-tipped catheter. N Engl J Med 283:447–451 - PubMed
-
- {'text': '', 'ref_index': 1, 'ids': [{'type': 'DOI', 'value': '10.1016/S0140-6736(05)67061-4', 'is_inner': False, 'url': 'https://doi.org/10.1016/s0140-6736(05)67061-4'}, {'type': 'PubMed', 'value': '16084255', 'is_inner': True, 'url': 'https://pubmed.ncbi.nlm.nih.gov/16084255/'}]}
- Harvey S, Harrison DA, Singer M, Ashcroft J, Jones CM, Elbourne D, Brampton W, Williams D, Young D, Rowan K, PAC-Man study collaboration (2005) Assessment of the clinical effectiveness of pulmonary artery catheters in management of patients in intensive care (PAC-Man): a randomized controlled trial. Lancet 366:472–477 - PubMed
-
- {'text': '', 'ref_index': 1, 'ids': [{'type': 'DOI', 'value': '10.1001/jama.290.20.2713', 'is_inner': False, 'url': 'https://doi.org/10.1001/jama.290.20.2713'}, {'type': 'PubMed', 'value': '14645314', 'is_inner': True, 'url': 'https://pubmed.ncbi.nlm.nih.gov/14645314/'}]}
- Richard C, Warszawski J, Anguel N, Deye N, Combes A, Barnoud D, Boulain T, Lefort Y, Fartoukh M, Baud F, Boyer A, Brochard L, Teboul JL, French Pulmonary Artery Catheter Study Group (2003) Early use of the pulmonary artery catheter and outcomes in patients with shock and acute respiratory distress syndrome: a randomized controlled trial. JAMA 290:2713–2720 - PubMed
-
- {'text': '', 'ref_index': 1, 'ids': [{'type': 'DOI', 'value': '10.1016/j.jcrc.2003.08.006', 'is_inner': False, 'url': 'https://doi.org/10.1016/j.jcrc.2003.08.006'}, {'type': 'PubMed', 'value': '14595569', 'is_inner': True, 'url': 'https://pubmed.ncbi.nlm.nih.gov/14595569/'}]}
- Peters SG, Afessa B, Decker PA, Schroeder DR, Offord KP, Scott JP (2003) Increased risk associated with pulmonary artery catheterization in the medical intensive care unit. J Crit Care 18:166–171 - PubMed
-
- {'text': '', 'ref_index': 1, 'ids': [{'type': 'DOI', 'value': '10.1016/S0147-9563(05)80036-6', 'is_inner': False, 'url': 'https://doi.org/10.1016/s0147-9563(05)80036-6'}, {'type': 'PubMed', 'value': '7622392', 'is_inner': True, 'url': 'https://pubmed.ncbi.nlm.nih.gov/7622392/'}]}
- Jensen L, Yakimets J, Teo KK (1995) A review of impedance cardiography. Heart Lung 24:183–193 - PubMed
Publication types