Pullout strength of suture anchors in comparison with transosseous sutures for rotator cuff repair
- PMID: 18193198
- DOI: 10.1007/s00167-007-0460-3
Pullout strength of suture anchors in comparison with transosseous sutures for rotator cuff repair
Abstract
Suture anchors are increasingly gaining importance in rotator cuff surgery. This means they will be gradually replacing transosseous sutures. The purpose of this study was to compare the stability of transosseous sutures with different suture anchors with regard to their pullout strength depending on bone density. By means of bone densitometry (CT scans), two groups of human humeral head specimens were determined: a healthy and a osteopenic bone group. Following anchor systems were being tested: SPIRALOK 5.0 mm (resorbable, DePuy Mitek), Super Revo 5 mm (titanium, Linvatec), UltraSorb (resorbable, Linvatec) and the double U-sutures with Orthocord USP 2 (partly resorbable, DePuy Mitek) and Ethibond Excel 2 (non-resorbable, Ethicon). The suture anchors/double U-sutures were inserted in the greater tuberosity 12 times. An electromechanical testing machine was used for cyclic loading with power increasing in stages. We recorded the ultimate failure loads, the system displacements and the modes of failure. The suture anchors tended to bring about higher ultimate failure loads than the transosseous double U-sutures. This difference was significant in the comparison of the Ethibond suture and the SPIRALOK 5.0 mm-both in healthy and osteopenic bone. Both the suture materials and the SPIRALOK 5.0 mm showed a significant difference in pullout strength on either healthy or osteopenic bone; the titanium anchor SuperRevo 5 mm and the tilting anchor UltraSorb did not show any significant difference in healthy or osteopenic bone. There was no significant difference concerning system displacement (healthy and osteopenic bone) between the five anchor systems tested. The pullout strength of transosseous sutures is neither on healthy nor on osteopenic bone higher than that of suture anchors. Therefore, even osteopenic bone does not constitute a valid reason for the surgeon to perform open surgery by means of transosseous sutures. The choice of sutures in osteopenic bone is of little consequence anyway since it is mostly the bone itself which is the limiting factor.
Similar articles
-
Suture anchor fixation strength in osteopenic versus non-osteopenic bone for rotator cuff repair.Arch Orthop Trauma Surg. 2009 Mar;129(3):373-9. doi: 10.1007/s00402-008-0689-4. Epub 2008 Jul 8. Arch Orthop Trauma Surg. 2009. PMID: 18607610
-
Cyclic load testing of biodegradable suture anchors containing 2 high-strength sutures.Arthroscopy. 2007 Apr;23(4):355-60. doi: 10.1016/j.arthro.2006.12.009. Arthroscopy. 2007. PMID: 17418326
-
Biomechanical testing of a new knotless suture anchor compared with established anchors for rotator cuff repair.J Shoulder Elbow Surg. 2008 Jul-Aug;17(4):642-6. doi: 10.1016/j.jse.2007.11.017. Epub 2008 Apr 8. J Shoulder Elbow Surg. 2008. PMID: 18396417
-
Biomechanical Strength of Rotator Cuff Repairs: A Systematic Review and Meta-regression Analysis of Cadaveric Studies.Am J Sports Med. 2019 Jul;47(8):1984-1993. doi: 10.1177/0363546518780928. Epub 2018 Jul 5. Am J Sports Med. 2019. PMID: 29975549
-
Arthroscopic rotator cuff repairs: an anatomic and biomechanical rationale for different suture-anchor repair configurations.Arthroscopy. 2007 Jun;23(6):662-9. doi: 10.1016/j.arthro.2007.02.018. Arthroscopy. 2007. PMID: 17560482 Review.
Cited by
-
Stress distribution inside bone after suture anchor insertion: simulation using a three-dimensional finite element method.Knee Surg Sports Traumatol Arthrosc. 2013 Aug;21(8):1777-82. doi: 10.1007/s00167-012-2060-0. Epub 2012 May 24. Knee Surg Sports Traumatol Arthrosc. 2013. PMID: 22622782
-
A novel technique of tibial tuberosity fracture fixation with two knotless suture anchors in an adult: A case report and literature review.Acta Orthop Traumatol Turc. 2022 Nov;56(6):416-420. doi: 10.5152/j.aott.2022.22102. Acta Orthop Traumatol Turc. 2022. PMID: 36567547 Free PMC article. Review.
-
Suture locking of isolated internal locking knotless suture anchors is not affected by bone quality.Open Access J Sports Med. 2015 Jun 22;6:201-8. doi: 10.2147/OAJSM.S81096. eCollection 2015. Open Access J Sports Med. 2015. PMID: 26124683 Free PMC article.
-
Is Arthroscopic Transosseous Rotator Cuff Repair Strength Dependent on the Tunnel Angle?Orthop J Sports Med. 2019 Jun 6;7(6):2325967119848667. doi: 10.1177/2325967119848667. eCollection 2019 Jun. Orthop J Sports Med. 2019. PMID: 31218236 Free PMC article.
-
Clinical Outcomes of Hip Abductor Repair Using Transosseous Sutures Versus Suture Anchors: A Systematic Review and Meta-analysis.Orthop J Sports Med. 2025 Jan 1;13(1):23259671241290320. doi: 10.1177/23259671241290320. eCollection 2025 Jan. Orthop J Sports Med. 2025. PMID: 39758143 Free PMC article. Review.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources