Comparative Study
. 2008 Jun;246(6):875-9.
doi: 10.1007/s00417-007-0758-3. Epub 2008 Jan 12.

Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer over a wide IOP range

Affiliations
Comparative Study

Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer over a wide IOP range

S Munkwitz et al. Graefes Arch Clin Exp Ophthalmol. 2008 Jun.

Abstract

Background: The aim of this study was to compare the intraocular pressure (IOP) results measured by the iCare rebound tonometer with those obtained by the Goldmann applanation tonometer (GAT) over a wide range of IOP values. Furthermore, the comfort level of the iCare measurement was evaluated.

Method: The study included 75 eyes of 75 patients. The patients were divided into three groups (7-15 mmHg n = 25, 16-22 mmHg n = 25, 23-60 mmHg n = 25). The measurements were taken by two independent observers in a masked fashion. All patients were asked about discomfort during the iCare measurement. To establish the agreement between the two devices, a Bland-Altman analysis was performed.

Results: Overall, the 95% confidence interval of the differences between the two devices was -8.67 to 10.25 mmHg and in 62.7%, the iCare measurement was within +/-3 mmHg of the GAT measurements. The distribution of the differences in IOP was similar, from 7-22 mmHg. In the higher IOP range (23-60 mmHg), however, the deviation was almost twice as large. The measurement with the iCare tonometer was well tolerated; 100% of the patients denied any discomfort.

Conclusions: The iCare tonometer is a mobile alternative to GAT in a low to moderate IOP range, but our findings show a greater deviation than previously reported. In high IOP values, measurements with the iCare tonometer do not correlate well with GAT.

PubMed Disclaimer

References

    1. Graefes Arch Clin Exp Ophthalmol. 2004 Jan;242(1):3-7 - PubMed
    1. Exp Eye Res. 2001 Dec;73(6):781-5 - PubMed
    1. J Glaucoma. 2006 Jun;15(3):213-7 - PubMed
    1. Arch Ophthalmol. 1990 Dec;108(12):1709-12 - PubMed
    1. Ophthalmologica. 1966;151(5):604-14 - PubMed

Publication types

LinkOut - more resources