Validation of aberrometry-based relative peripheral refraction measurements
- PMID: 18201339
- DOI: 10.1111/j.1475-1313.2007.00535.x
Validation of aberrometry-based relative peripheral refraction measurements
Abstract
Purpose: To validate aberrometry-based relative peripheral refraction (RPR) measurements and to examine RPR differences in the nasal and temporal horizontal visual field.
Methods: Thirty subjects underwent cycloplegia in the right eye. Ten measurements each were made centrally, 30 degrees nasally on the retina and 30 degrees temporally on the retina using the Complete Ophthalmic Analysis System (COAS) aberrometer and the Grand Seiko WR-5100K autorefractor in random order. A 2-mm COAS analysis diameter was used to approximate the Grand Seiko autorefractor measurement beam diameter. Nasal and temporal RPR were calculated as the difference between the average nasal or temporal spherical equivalent and the average central spherical equivalent. Repeated measures ANOVAs were used to test for significant differences between the RPR readings of each instrument in each retinal location (nasal or temporal) and to examine refractive error in each direction of gaze for both instruments.
Results: Central spherical equivalent refractive error as measured by the Grand Seiko autorefractor ranged from +0.63 D to -8.41 D (mean +/- S.D. = -2.63 D +/- 2.05 D). There were no significant differences between the instruments for RPR measurements (p = 0.34). The nasal RPR was significantly more hyperopic than the temporal RPR (mean +/- S.D. = +0.45 D +/- 1.04 D, p = 0.02). Spherical equivalent refractive error in all directions of gaze was more myopic with the COAS than with the Grand Seiko autorefractor (mean +/- S.D. = -0.41 D +/- 0.61 D, p < 0.0001).
Conclusions: RPR measurements with the COAS are feasible and equivalent to those made with the Grand Seiko autorefractor. The COAS can be used to simultaneously collect RPR and peripheral aberration data. Nasal and temporal RPR measurements were significantly different in our sample.
Similar articles
-
Systematic measurement errors involved in over-refraction using an autorefractor (Grand-Seiko WV-500): is measurement of accommodative lag through spectacle lenses valid?Ophthalmic Physiol Opt. 2007 May;27(3):281-6. doi: 10.1111/j.1475-1313.2007.00466.x. Ophthalmic Physiol Opt. 2007. PMID: 17470241
-
Experimental investigation of accommodation in eyes fit with multifocal contact lenses using a clinical auto-refractor.Ophthalmic Physiol Opt. 2018 Mar;38(2):152-163. doi: 10.1111/opo.12438. Epub 2018 Jan 8. Ophthalmic Physiol Opt. 2018. PMID: 29315718
-
Evaluation of a clinical aberrometer for lower-order accuracy and repeatability, higher-order repeatability, and instrument myopia.Optometry. 2005 Aug;76(8):461-72. doi: 10.1016/j.optm.2005.07.006. Optometry. 2005. PMID: 16150413
-
A review of peripheral refraction techniques.Optom Vis Sci. 2009 May;86(5):429-46. doi: 10.1097/OPX.0b013e31819fa727. Optom Vis Sci. 2009. PMID: 19342977 Review.
-
[Aberrometer systems in laser refractive surgery].Klin Oczna. 2005;107(10-12):719-21. Klin Oczna. 2005. PMID: 16619829 Review. Polish.
Cited by
-
New Objective Refraction Metric Based on Sphere Fitting to the Wavefront.J Ophthalmol. 2017;2017:1909348. doi: 10.1155/2017/1909348. Epub 2017 Sep 20. J Ophthalmol. 2017. PMID: 29104804 Free PMC article.
-
Central and peripheral autorefraction repeatability in normal eyes.Optom Vis Sci. 2014 Sep;91(9):1106-12. doi: 10.1097/OPX.0000000000000351. Optom Vis Sci. 2014. PMID: 25062133 Free PMC article.
-
Experimental and modeling analysis of lenses with concentric cylindrical annular refractive elements: impact on peripheral imaging.Biomed Opt Express. 2025 Mar 7;16(4):1344-1358. doi: 10.1364/BOE.546942. eCollection 2025 Apr 1. Biomed Opt Express. 2025. PMID: 40321993 Free PMC article.
-
Aberrometry Repeatability and Agreement with Autorefraction.Optom Vis Sci. 2017 Sep;94(9):886-893. doi: 10.1097/OPX.0000000000001107. Optom Vis Sci. 2017. PMID: 28727613 Free PMC article.
-
Comparison of an open view autorefractor with an open view aberrometer in determining peripheral refraction in children.J Optom. 2023 Jan-Mar;16(1):20-29. doi: 10.1016/j.optom.2021.12.002. Epub 2022 Jan 10. J Optom. 2023. PMID: 35022155 Free PMC article.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Medical