Comparative Study
. 2008 Jan;15(1):66-73.
doi: 10.1111/j.1553-2712.2007.00007.x.

TASER X26 discharges in swine produce potentially fatal ventricular arrhythmias

Affiliations
Free article
Comparative Study

TASER X26 discharges in swine produce potentially fatal ventricular arrhythmias

Robert J Walter et al. Acad Emerg Med. 2008 Jan.
Free article

Abstract

Objectives: Data from the authors and others suggest that TASER X26 stun devices can acutely alter cardiac function in swine. The authors hypothesized that TASER discharges degrade cardiac performance through a mechanism not involving concurrent acidosis.

Methods: Using an Institutional Animal Care and Use Committee (IACUC)-approved protocol, Yorkshire pigs (25-71 kg) were anesthetized, paralyzed with succinylcholine (SCh; 2 mg/kg), and then exposed to two 40-second discharges from a TASER X26 with a transcardiac vector. Vital signs, blood chemistry, and electrolyte levels were obtained before exposure and periodically for 48 hours postdischarge. Electrocardiograms and echocardiography (echo) were performed before, during, and after the discharges. p-Values < 0.05 were considered significant.

Results: Electrocardiograms were unreadable during the discharges due to electrical interference, but echo images showed unmistakably that cardiac rhythm was captured immediately at a rate of 301 +/- 18 beats/min (n = 8) in all animals tested. Capture continued for the duration of the discharge and in one animal degenerated into fatal ventricular fibrillation (VF). In the remaining animals, ventricular tachycardia (VT) occurred postdischarge for 1-17 seconds, whereupon sinus rhythm was regained spontaneously. Blood chemistry values and vital signs were minimally altered postdischarge and no significant acidosis was seen.

Conclusions: Extreme acid-base disturbances usually seen after lengthy TASER discharges were absent with SCh, but TASER X26 discharges immediately and invariably produced myocardial capture. This usually reverted spontaneously to sinus rhythm postdischarge, but fatal VF was seen in one animal. Thus, in the absence of systemic acidosis, lengthy transcardiac TASER X26 discharges (2 x 40 seconds) captured myocardial rhythm, potentially resulting in VT or VF in swine.

PubMed Disclaimer

Comment in

  • Sensitive swine and TASER electronic control devices.
    Kroll MW, Calkins H, Luceri RM, Graham MA, Heegaard WG. Kroll MW, et al. Acad Emerg Med. 2008 Jul;15(7):695-6; author reply 696-8. doi: 10.1111/j.1553-2712.2008.00141.x. Acad Emerg Med. 2008. PMID: 19086217 No abstract available.