Gas bubble motion artifact in MDCT
- PMID: 18212212
- DOI: 10.2214/AJR.07.2702
Gas bubble motion artifact in MDCT
Abstract
Objective: The purpose of our study was to characterize the imaging features of an MDCT artifact caused by gas bubble motion.
Materials and methods: For the period 2002-2006, the CT images of 10 patients that revealed a curvilinear artifact thought to be due to a gas bubble moving during CT acquisition were retrospectively identified and reviewed by an attending body radiologist. The clinical images were acquired on MDCT scanners. A phantom containing water and moving air bubbles (gas injection rates of 0.1 and 0.5 mL/s) was designed, built, and scanned using a 16-MDCT scanner. Computer simulation was used to test our hypothesis that these observed artifacts originated from moving gas bubbles.
Results: Semicircular tubular CT artifacts with attenuation close to that of air were observed in all 10 clinical cases. The acquired MDCT images from the phantom and the computer simulations showed tubular air-attenuation semicircular artifacts that closely matched the imaging findings on the 10 clinical cases. Increased rates of bubble injection multiplied the artifact. Based on the computer simulation, the precise appearance of the artifact depends on the bubble location and velocity relative to the rotation of the CT scanner.
Conclusion: Gas bubble motion during CT generates a semicircular air-attenuation artifact that has not been previously described to our knowledge. The artifact is typically found in bowel and other liquid-containing structures in which a bubble of gas floats up through a liquid medium during CT.
Comment in
-
Gas bubble motion artifact.AJR Am J Roentgenol. 2008 Dec;191(6):W312. doi: 10.2214/AJR.08.1055. AJR Am J Roentgenol. 2008. PMID: 19020221 No abstract available.
Similar articles
-
Gas bubble motion artifact.AJR Am J Roentgenol. 2008 Dec;191(6):W312. doi: 10.2214/AJR.08.1055. AJR Am J Roentgenol. 2008. PMID: 19020221 No abstract available.
-
Comparison of artifacts on coronal reformation and axial CT pulmonary angiography images using single-detector and 4- and 8-detector multidetector-row helical CT scanners.Acad Radiol. 2005 May;12(5):602-7. doi: 10.1016/j.acra.2005.01.011. Acad Radiol. 2005. PMID: 15866133
-
Thin-section CT of lung without ECG gating: 64-detector row CT can markedly reduce cardiac motion artifact which can simulate lung lesions.Eur J Radiol. 2009 Jan;69(1):102-7. doi: 10.1016/j.ejrad.2007.10.027. Epub 2007 Dec 19. Eur J Radiol. 2009. PMID: 18068924
-
X-ray-based attenuation correction for positron emission tomography/computed tomography scanners.Semin Nucl Med. 2003 Jul;33(3):166-79. doi: 10.1053/snuc.2003.127307. Semin Nucl Med. 2003. PMID: 12931319 Review.
-
Sonography of Gastrointestinal Tract Diseases: Correlation With Computed Tomographic Findings and Endoscopy.J Ultrasound Med. 2016 Jul;35(7):1543-71. doi: 10.7863/ultra.15.09038. Epub 2016 Jun 7. J Ultrasound Med. 2016. PMID: 27268998 Review.
Cited by
-
Reduction of peristalsis-related gastrointestinal streak artifacts with dual-energy CT: a patient and phantom study.Abdom Radiol (NY). 2016 Aug;41(8):1456-65. doi: 10.1007/s00261-016-0702-2. Abdom Radiol (NY). 2016. PMID: 26987848 Free PMC article.
-
Four-dimensional computed tomography of the left ventricle, Part I: Motion artifact reduction.Med Phys. 2022 Jul;49(7):4404-4418. doi: 10.1002/mp.15709. Epub 2022 May 30. Med Phys. 2022. PMID: 35588288 Free PMC article.
-
Motion Induced Artifact Mimicking Cervical Dens Fracture on the CT Scan: A Case Report.Asian Spine J. 2012 Sep;6(3):216-8. doi: 10.4184/asj.2012.6.3.216. Epub 2012 Aug 21. Asian Spine J. 2012. PMID: 22977704 Free PMC article.
-
Reducing Visceral-Motion-Related Artifacts on the Liver with Dual-Energy CT: A Comparison of Four Different CT Scanner Techniques.Diagnostics (Basel). 2022 Sep 5;12(9):2155. doi: 10.3390/diagnostics12092155. Diagnostics (Basel). 2022. PMID: 36140556 Free PMC article.
-
Reduction of Peristalsis-Related Streak Artifacts on the Liver with Dual-Layer Spectral CT.Diagnostics (Basel). 2022 Mar 23;12(4):782. doi: 10.3390/diagnostics12040782. Diagnostics (Basel). 2022. PMID: 35453830 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical