. 2008 May;30(5):618-21.
doi: 10.1002/hed.20764.

Investigation of tracheoesophageal voice prosthesis leakage patterns: patient's self-report versus clinician's confirmation

Affiliations

Investigation of tracheoesophageal voice prosthesis leakage patterns: patient's self-report versus clinician's confirmation

Lynn M Acton et al. Head Neck. 2008 May.

Erratum in

  • Head Neck. 2008 Dec;30(12):1663

Abstract

Background: This study investigated the patient's self-report and clinician's confirmation of tracheoesophageal voice prosthesis leakage patterns (through or around) with or without the cough reflex and whether prosthesis diameter affected the leakage route.

Methods: Sixty-six consecutive participants with a total of 200 patient-initiated reasons for prosthesis changes were enrolled prospectively. Patient's self-report of leakage and cough reflex were recorded prior to clinician's confirmation.

Results: One-hundred eight (54%) of the 200 patient-initiated reasons for prosthesis changes were leakage through or around the voice prosthesis. Leakage was unrecognized in 21 (23%) of 92 instances, even though 15 (71%) of those 21 instances exhibited a cough reflex. Clinician's confirmed leakage in 118 (59%) of 200 patient-initiated reasons for prosthesis changes. Coughing occurred significantly less with leakage around (9 [53%] of 17 instances) than that with leakage through the voice prosthesis (80 [88%] of 91 instances) (chi(2) [1, N=108], p< .05). Leakage around the voice prosthesis occurred more with 20-Fr diameter prostheses (16 [76%] of 21 instances).

Conclusions: Patient education is important for reliable identification of leakage for prompt prosthesis replacement. Leakage around the voice prosthesis can be minimized or avoided by initially fitting and continuing the use of smaller diameter (16 Fr) voice prostheses.