Reader studies for validation of CAD systems
- PMID: 18215501
- DOI: 10.1016/j.neunet.2007.12.013
Reader studies for validation of CAD systems
Erratum in
- Neural Netw. 2008 May;21(4):698
Abstract
Evaluation of computational intelligence (CI) systems designed to improve the performance of a human operator is complicated by the need to include the effect of human variability. In this paper we consider human (reader) variability in the context of medical imaging computer-assisted diagnosis (CAD) systems, and we outline how to compare the detection performance of readers with and without the CAD. An effective and statistically powerful comparison can be accomplished with a receiver operating characteristic (ROC) experiment, summarized by the reader-averaged area under the ROC curve (AUC). The comparison requires sophisticated yet well-developed methods for multi-reader multi-case (MRMC) variance analysis. MRMC variance analysis accounts for random readers, random cases, and correlations in the experiment. In this paper, we extend the methods available for estimating this variability. Specifically, we present a method that can treat arbitrary study designs. Most methods treat only the fully-crossed study design, where every reader reads every case in two experimental conditions. We demonstrate our method with a computer simulation, and we assess the statistical power of a variety of study designs.
Similar articles
-
One-shot estimate of MRMC variance: AUC.Acad Radiol. 2006 Mar;13(3):353-62. doi: 10.1016/j.acra.2005.11.030. Acad Radiol. 2006. PMID: 16488848
-
Breast MRI lesion classification: improved performance of human readers with a backpropagation neural network computer-aided diagnosis (CAD) system.J Magn Reson Imaging. 2007 Jan;25(1):89-95. doi: 10.1002/jmri.20794. J Magn Reson Imaging. 2007. PMID: 17154399
-
Hierarchical models for ROC curve summary measures: design and analysis of multi-reader, multi-modality studies of medical tests.Stat Med. 2008 Jan 30;27(2):243-56. doi: 10.1002/sim.2828. Stat Med. 2008. PMID: 17340598
-
Assessment of medical imaging systems and computer aids: a tutorial review.Acad Radiol. 2007 Jun;14(6):723-48. doi: 10.1016/j.acra.2007.03.001. Acad Radiol. 2007. PMID: 17502262 Review.
-
The use of receiver operating characteristic curves in biomedical informatics.J Biomed Inform. 2005 Oct;38(5):404-15. doi: 10.1016/j.jbi.2005.02.008. Epub 2005 Apr 2. J Biomed Inform. 2005. PMID: 16198999 Review.
Cited by
-
Relationship between Obuchowski-Rockette-Hillis and Gallas methods for analyzing multi-reader diagnostic imaging data with empirical AUC as the reader performance measure.Biostat Epidemiol. 2023;7(1):e2062115. doi: 10.1080/24709360.2022.2062115. Epub 2022 Jun 10. Biostat Epidemiol. 2023. PMID: 37719421 Free PMC article.
-
Evaluating imaging and computer-aided detection and diagnosis devices at the FDA.Acad Radiol. 2012 Apr;19(4):463-77. doi: 10.1016/j.acra.2011.12.016. Epub 2012 Feb 3. Acad Radiol. 2012. PMID: 22306064 Free PMC article.
-
Greater benefits of deep learning-based computer-aided detection systems for finding small signals in 3D volumetric medical images.J Med Imaging (Bellingham). 2024 Jul;11(4):045501. doi: 10.1117/1.JMI.11.4.045501. Epub 2024 Jul 9. J Med Imaging (Bellingham). 2024. PMID: 38988989 Free PMC article.
-
Evaluation of computer-aided detection and diagnosis systems.Med Phys. 2013 Aug;40(8):087001. doi: 10.1118/1.4816310. Med Phys. 2013. PMID: 23927365 Free PMC article.
-
Impact of prevalence and case distribution in lab-based diagnostic imaging studies.J Med Imaging (Bellingham). 2019 Jan;6(1):015501. doi: 10.1117/1.JMI.6.1.015501. Epub 2019 Jan 21. J Med Imaging (Bellingham). 2019. PMID: 30713851 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous