The effects of peroxide content on the wear behavior, microstructure and mechanical properties of peroxide crosslinked ultra-high molecular weight polyethylene used in total hip replacement
- PMID: 18219557
- DOI: 10.1007/s10856-008-3368-7
The effects of peroxide content on the wear behavior, microstructure and mechanical properties of peroxide crosslinked ultra-high molecular weight polyethylene used in total hip replacement
Abstract
The wear of the ultra-high molecular weight polyethylene (UHMWPE) acetabular components and wear debris induced osteolysis are the major causes of failure in total hip replacements. Crosslinking has been shown to improve the wear resistance of UHMWPE by producing a network structure, resisting the plastic deformation of the surface layer. In this study organic peroxides were used to crosslink two different types of UHMWPE resins, using hot isostatic pressing as the processing method. The effects of peroxide content on the different properties were investigated, along with the effect of the crosslink density on the wear behavior. An increase in peroxide content decreases the melting point and the degree of crystallinity, which results in a decrease in the yield strength. The ultimate tensile strength remains essentially unchanged. The molecular weight between crosslinks decreases with an increase in the peroxide content and reaches a saturation limit at around 0.3-0.5 weight percent peroxide, its value at the saturation limit is a function of the virgin resin used for processing. The wear rate decreases linearly with the increase in crosslink density.
Similar articles
-
Unified wear model for highly crosslinked ultra-high molecular weight polyethylenes (UHMWPE).Biomaterials. 1999 Aug;20(16):1463-70. doi: 10.1016/s0142-9612(99)00039-3. Biomaterials. 1999. PMID: 10458559
-
Development of an extremely wear-resistant ultra high molecular weight polyethylene for total hip replacements.J Orthop Res. 1999 Mar;17(2):157-67. doi: 10.1002/jor.1100170203. J Orthop Res. 1999. PMID: 10221831
-
Surface cross-linked UHMWPE using peroxides.J Orthop Res. 2017 Nov;35(11):2551-2556. doi: 10.1002/jor.23569. Epub 2017 Jun 2. J Orthop Res. 2017. PMID: 28374917
-
Wear products of total hip arthroplasty: The case of polyethylene.Morphologie. 2017 Mar;101(332):1-8. doi: 10.1016/j.morpho.2016.06.001. Epub 2016 Jul 15. Morphologie. 2017. PMID: 27426252 Review.
-
Implant wear mechanisms--basic approach.Biomed Mater. 2008 Dec;3(4):042001. doi: 10.1088/1748-6041/3/4/042001. Epub 2008 Sep 25. Biomed Mater. 2008. PMID: 18824778 Review.