Comparison of dynamic contour tonometry and goldmann applanation tonometry in deep lamellar and penetrating keratoplasties
- PMID: 18222191
- DOI: 10.1016/j.ajo.2007.10.003
Comparison of dynamic contour tonometry and goldmann applanation tonometry in deep lamellar and penetrating keratoplasties
Abstract
Purpose: To compare the measurements of intraocular pressure (IOP) with dynamic contour tonometry (DCT) and Goldmann applanation tonometry (GAT) in eyes with corneal graft and to evaluate the influence of corneal thickness (CCT), corneal curvature (CC), and astigmatism on these methods.
Design: Prospective, observational cross-sectional study.
Methods: Eighteen eyes of 18 patients after penetrating keratoplasty (PKP) and 14 eyes of 14 patients after deep lamellar keratoplasty (DLKP) underwent IOP evaluation with DCT and GAT, and measurements of CCT, CC, and astigmatism. Bland-Altman plots were used to evaluate the agreement between tonometers. Multivariate regression analysis was used to evaluate the influence of ocular structural factors and running suture on IOP measurements obtained with both tonometers.
Results: IOP values obtained by DCT and GAT were strongly correlated in all eyes (r = .91; P < .001). DCT values measured 2.5 +/- 1.7 mm Hg higher than GAT readings (P < .001). A reduction of the mean IOP difference between DCT and GAT with an increase in IOP values (P < .001) was found. Regression analysis showed no effect of CCT, CC, astigmatism, and running suture on both DCT and GAT readings, either in DLKP or in PKP eyes.
Conclusions: We found a good overall correlation between both tonometers but the agreement between instruments differs in high or low IOP ranges. The wide and varying 95% limits of agreement between DCT and GAT indicates that DCT provides IOP measurements on deep lamellar and penetrating keratoplasties which can be used in the clinical practice.
Similar articles
-
Comparison of Goldmann applanation tonometry and dynamic contour tonometry in healthy and glaucomatous eyes.Eye (Lond). 2009 Feb;23(2):262-9. doi: 10.1038/sj.eye.6703102. Epub 2008 Jan 25. Eye (Lond). 2009. PMID: 18219335
-
Comparison of dynamic contour tonometry with goldmann applanation tonometry.Invest Ophthalmol Vis Sci. 2004 Sep;45(9):3118-21. doi: 10.1167/iovs.04-0018. Invest Ophthalmol Vis Sci. 2004. PMID: 15326129 Clinical Trial.
-
Comparison of dynamic contour tonometry with Goldmann applanation tonometry following Descemet's stripping automated endothelial keratoplasty (DSAEK).Klin Monbl Augenheilkd. 2009 Apr;226(4):241-4. doi: 10.1055/s-0028-1109242. Epub 2009 Apr 21. Klin Monbl Augenheilkd. 2009. PMID: 19384775 Clinical Trial.
-
[What to do if the intraocular pressure measurement does not appear reliable].J Fr Ophtalmol. 2010 Apr;33(4):279-84. doi: 10.1016/j.jfo.2010.02.007. Epub 2010 Mar 29. J Fr Ophtalmol. 2010. PMID: 20347507 Review. French.
-
New ways to measure intraocular pressure.Curr Opin Ophthalmol. 2008 Mar;19(2):122-6. doi: 10.1097/ICU.0b013e3282f391ae. Curr Opin Ophthalmol. 2008. PMID: 18301285 Review.
Cited by
-
Comparison of Intraocular Pressure Measurements Using Three Different Methods (Goldmann Applanation Tonometry (GAT), Corvis ST, and iCare) Following Penetrating Keratoplasty.J Clin Med. 2024 Nov 22;13(23):7046. doi: 10.3390/jcm13237046. J Clin Med. 2024. PMID: 39685505 Free PMC article.
-
Repeatability of intraocular pressure measurements with Icare PRO rebound, Tono-Pen AVIA, and Goldmann tonometers in sitting and reclining positions.BMC Ophthalmol. 2013 Sep 5;13:44. doi: 10.1186/1471-2415-13-44. BMC Ophthalmol. 2013. PMID: 24006952 Free PMC article. Clinical Trial.
-
Challenges in the complex management of post-keratoplasty glaucoma.Ther Adv Ophthalmol. 2021 Jul 20;13:25158414211031397. doi: 10.1177/25158414211031397. eCollection 2021 Jan-Dec. Ther Adv Ophthalmol. 2021. PMID: 34350382 Free PMC article. Review.
-
Glaucoma after penetrating keratoplasty.Rom J Ophthalmol. 2017 Jul-Sep;61(3):159-165. doi: 10.22336/rjo.2017.30. Rom J Ophthalmol. 2017. PMID: 29450392 Free PMC article. Review.
-
Comparison of iCare tonometer and Goldmann applanation tonometry in normal corneas and in eyes with automated lamellar and penetrating keratoplasty.Eye (Lond). 2011 May;25(5):642-50. doi: 10.1038/eye.2011.60. Epub 2011 Mar 25. Eye (Lond). 2011. PMID: 21436848 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources