Blood pressure changes associated with tilting in normotensive subjects: differences in response pattern as measured by oscillometry and auscultation
- PMID: 1822764
- DOI: 10.1007/BF01826214
Blood pressure changes associated with tilting in normotensive subjects: differences in response pattern as measured by oscillometry and auscultation
Abstract
Three non-invasive instruments were used to measure blood pressure in the supine position and on tilting--a conventional and a random-zero sphygmomanometer, and an oscillometric device (Accutorr 1A). Twenty normotensives volunteered for the study. There was no statistically significant difference in systolic blood pressure and diastolic blood pressure measured by the conventional and random-zero sphygmomanometers in the supine position. There was a difference between these recordings and those of the Accutorr, with the Accutorr giving higher readings of systolic blood pressure (p less than 0.001), analysis of variance, 95% confidence interval of the difference between the Accutorr and the random-zero was 5.1-15.7 mmHg) and lower readings of diastolic blood pressure (p less than 0.0001, analysis of variance, 95% confidence interval of the difference between the Accutorr and the random-zero was -12.2- -2.2 mmHg). On tilting, the Accutorr showed an increase in systolic blood pressure while the other two machines did not (p less than 0.01, analysis of variance). By contrast, the Accutorr detected a smaller rise in diastolic blood pressure than with the other two instruments (p less than 0.05, analysis of variance). The difference between blood pressure measurements made in the supine position by the two different techniques, auscultation and oscillometry, might be expected. However, the two different techniques do not detect the same blood pressure responses to a change in posture.
Similar articles
-
Comparison of blood pressure measurements between an automated oscillometric device and a Hawksley random-zero sphygmomanometer in the northern Sweden MONICA study.Blood Press Monit. 2012 Aug;17(4):164-70. doi: 10.1097/MBP.0b013e328356ef58. Blood Press Monit. 2012. PMID: 22781634 Clinical Trial.
-
Validation of three oscillometric blood pressure devices against auscultatory mercury sphygmomanometer in children.Blood Press Monit. 2006 Oct;11(5):281-91. doi: 10.1097/01.mbp.0000209082.09623.b4. Blood Press Monit. 2006. PMID: 16932037 Clinical Trial.
-
Postural effects on diastolic blood pressure are differently recorded by a non-invasive ambulatory blood pressure monitor and a standard auscultatory device.Clin Physiol. 1993 Nov;13(6):621-9. doi: 10.1111/j.1475-097x.1993.tb00477.x. Clin Physiol. 1993. PMID: 8119056
-
Comparison of oscillometric blood pressure measurements at the wrist with an upper-arm auscultatory mercury sphygmomanometer.Clin Exp Pharmacol Physiol. 1999 May-Jun;26(5-6):477-81. doi: 10.1046/j.1440-1681.1999.03059.x. Clin Exp Pharmacol Physiol. 1999. PMID: 10386242
-
Arterial stiffness as underlying mechanism of disagreement between an oscillometric blood pressure monitor and a sphygmomanometer.Hypertension. 2000 Oct;36(4):484-8. doi: 10.1161/01.hyp.36.4.484. Hypertension. 2000. PMID: 11040223 Clinical Trial.
Cited by
-
Yohimbine in neurally mediated syncope. Pathophysiological implications.J Clin Invest. 1998 Nov 15;102(10):1824-30. doi: 10.1172/JCI3050. J Clin Invest. 1998. PMID: 9819368 Free PMC article. Clinical Trial.
References
Publication types
MeSH terms
LinkOut - more resources
Medical