Is Doppler planimetry a valid technique for the evaluation of postpartum urinary bladder volume?
Abstract
The objective of this prospective study was to evaluate the accuracy of conventional 2D ultrasound (CUS) versus doppler planimetry (DP) in the assessment of postpartum urinary bladder volume compared to a true estimate using urethral catheterisation. Fifty-two women were assessed within 24 hours of delivery. Evaluation of bladder volume was performed using CUS (1-estimate) and DP (6-estimates). CUS had a higher correlation (r=0.796) with the true volume and lower % error than DP in the postpartum group. DP readings were highly reproducible (ICC 0.81) but tended to overestimate the true value especially with smaller volumes. DP was suboptimal for the assessment of the postpartum PVR. Postpartum evaluation using CUS is more accurate in calculating the true urinary volume.
Similar articles
-
Is portable three-dimensional ultrasound a valid technique for measurement of postpartum urinary bladder volume?Taiwan J Obstet Gynecol. 2014 Mar;53(1):12-6. doi: 10.1016/j.tjog.2013.01.028. Taiwan J Obstet Gynecol. 2014. PMID: 24767639 Clinical Trial.
-
Comparative accuracy of the Lilium α-200 portable ultrasound bladder scanner and conventional transabdominal ultrasonography for postvoid residual urine volume measurement in association with the clinical factors involved in measurement errors.Neurourol Urodyn. 2021 Jan;40(1):183-192. doi: 10.1002/nau.24530. Epub 2020 Oct 6. Neurourol Urodyn. 2021. PMID: 33022795
-
Accuracy of bladder volume determinations by ultrasonography: are they accurate over entire bladder volume range?Urology. 2003 Oct;62(4):656-60. doi: 10.1016/s0090-4295(03)00559-4. Urology. 2003. PMID: 14550437
-
Ultrasound measurement of fetal urine flow.Clin Obstet Gynecol. 1997 Jun;40(2):337-51. doi: 10.1097/00003081-199706000-00010. Clin Obstet Gynecol. 1997. PMID: 9199845 Review.
-
Modifications of the bladder wall (organ damage) in patients with bladder outlet obstruction: ultrasound parameters.Arch Ital Urol Androl. 2012 Dec;84(4):263-7. Arch Ital Urol Androl. 2012. PMID: 23427760 Review.