5-year experience of highly cross-linked polyethylene in cemented and uncemented sockets: two randomized studies using radiostereometric analysis
- PMID: 18236180
- DOI: 10.1080/17453670710014518
5-year experience of highly cross-linked polyethylene in cemented and uncemented sockets: two randomized studies using radiostereometric analysis
Abstract
Background: Laboratory tests and early clinical studies have shown that highly cross-linked polyethylene (PE) markedly improves wear resistance compared to conventional PE. We evaluated this type of PE in two randomized clinical studies using radiostereometric analysis (RSA). The 2- and 3-year follow-up of these studies have already been reported. We found a lower penetration rate for the highly cross-linked PE than for conventional PE. We now report the outcome after 5 years.
Patients and methods: 60 patients (61 hips) with a median age of 55 years were randomized to receive either highly cross-linked PE (Durasul) or conventional cemented all-polyethylene sockets of the same design. 55 patients (56 hips) were followed for 5 years. In the second study, 32 patients (64 hips) with a median age of 48 years and with bilateral primary or secondary osteoarthritis of the hip had hybrid total hip arthroplasty with liners made of highly cross-linked PE (Longevity) on one side and conventional PE on the other. 23 patients in this study have passed the 5-year follow-up.
Results: Bedding-in and creep for the Durasul all-PE was reached by 2 years, and by 1 year for the Longevity liners. In both control groups with conventional PE, this was reached by 6 months. The steady-state wear rate was close to zero in the two study groups and 0.06 mm/year for the conventional PE in the two control groups.
Interpretation: We found that use of highly crosslinked PE instead of conventional PE reduced the wear by more than 95%, which supports continued use of this type of PE in young, active patients.
Similar articles
-
The Otto Aufranc Award. Highly cross-linked polyethylene in total hip arthroplasty: randomized evaluation of penetration rate in cemented and uncemented sockets using radiostereometric analysis.Clin Orthop Relat Res. 2004 Dec;(429):6-16. Clin Orthop Relat Res. 2004. PMID: 15577460 Review.
-
The seven-year wear of highly cross-linked polyethylene in total hip arthroplasty: a double-blind, randomized controlled trial using radiostereometric analysis.J Bone Joint Surg Am. 2011 Apr 20;93(8):716-22. doi: 10.2106/JBJS.J.00287. J Bone Joint Surg Am. 2011. PMID: 21508278 Clinical Trial.
-
New polymer materials in total hip arthroplasty. Evaluation with radiostereometry, bone densitometry, radiography and clinical parameters.Acta Orthop Suppl. 2005 Feb;76(315):3-82. Acta Orthop Suppl. 2005. PMID: 15790289 Review.
-
Differences in Femoral Head Penetration Between Highly Cross-Linked Polyethylene Cemented Sockets and Uncemented Liners.J Arthroplasty. 2017 Dec;32(12):3796-3801. doi: 10.1016/j.arth.2017.07.017. Epub 2017 Jul 21. J Arthroplasty. 2017. PMID: 28781017
-
The creep and wear of highly cross-linked polyethylene: a three-year randomised, controlled trial using radiostereometric analysis.J Bone Joint Surg Br. 2008 May;90(5):556-61. doi: 10.1302/0301-620X.90B5.20545. J Bone Joint Surg Br. 2008. PMID: 18450618 Clinical Trial.
Cited by
-
Is the use of thin, highly cross-linked polyethylene liners safe in total hip arthroplasty?Int Orthop. 2016 Apr;40(4):681-6. doi: 10.1007/s00264-015-2841-4. Epub 2015 Jul 2. Int Orthop. 2016. PMID: 26130285
-
Wear of a sequentially annealed polyethylene acetabular liner.Acta Orthop. 2014 Sep;85(5):470-3. doi: 10.3109/17453674.2014.949044. Epub 2014 Aug 20. Acta Orthop. 2014. PMID: 25140986 Free PMC article. Clinical Trial.
-
Polyethylene wear and cup migration of cemented total hip arthroplasty with femoral heads made of oxidized zirconium, steel, or cobalt chromium: a 10-year secondary analysis from a randomized trial using radiostereometry.Acta Orthop. 2024 Sep 30;95:578-585. doi: 10.2340/17453674.2024.41945. Acta Orthop. 2024. PMID: 39347798 Free PMC article. Clinical Trial.
-
Biomechanical concept and clinical outcome of dual mobility cups.Int Orthop. 2012 Dec;36(12):2411-8. doi: 10.1007/s00264-012-1678-3. Epub 2012 Oct 17. Int Orthop. 2012. PMID: 23073926 Free PMC article. Review.
-
Dislocation after total hip arthroplasty with 28 and 32-mm femoral head.J Orthop Traumatol. 2010 Jun;11(2):111-5. doi: 10.1007/s10195-010-0097-8. Epub 2010 May 27. J Orthop Traumatol. 2010. PMID: 20505973 Free PMC article.