Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Randomized Controlled Trial
. 2008 Apr;100(4):544-8.
doi: 10.1093/bja/aen002. Epub 2008 Jan 31.

Pentax-AWS, a new videolaryngoscope, is more effective than the Macintosh laryngoscope for tracheal intubation in patients with restricted neck movements: a randomized comparative study

Affiliations
Free article
Randomized Controlled Trial

Pentax-AWS, a new videolaryngoscope, is more effective than the Macintosh laryngoscope for tracheal intubation in patients with restricted neck movements: a randomized comparative study

Y Enomoto et al. Br J Anaesth. 2008 Apr.
Free article

Abstract

Background: We studied whether laryngoscopy and tracheal intubation were easier when using the Pentax-AWS (Tokyo, Japan), a new videolaryngoscope, than when applying the Macintosh laryngoscope, during manual in-line neck stabilization.

Methods: In 203 anaesthetized patients with manual in-line neck stabilization, we inserted the Pentax-AWS and a Macintosh laryngoscope, in turn, and recorded the view of the glottis and time taken to laryngoscopy. The success rate of tracheal intubation (within 120 s) and time to intubation were also recorded.

Results: The view of the glottis was significantly better with the Pentax-AWS than with the Macintosh laryngoscope (P<0.001). For the Macintosh laryngoscope, the view was obscured in 22 of 203 patients (11%) (Grade 3 in 21 patients and Grade 4 in one patient), whereas for the Pentax-AWS, the glottis was always clearly seen (Grade 1). Time taken to see the glottis with the Pentax-AWS [mean (sd): 6.0 (3.1) s] was significantly shorter than with the Macintosh laryngoscope [11.0 (5.0) s] (95% CI for difference: 4-6 s). The success rate of tracheal intubation with the Pentax-AWS (all of 99 patients) was significantly higher than with the Macintosh laryngoscope (93 of 104 patients) (P=0.001). Time taken for intubation was similar between the Macintosh laryngoscope [51 (27) s] and the Pentax-AWS [54 (14) s] (95% CI for difference: -9 to 3 s).

Conclusions: In patients with stabilized neck, the Pentax-AWS provided a better view of the glottis and a higher success rate of tracheal intubation, compared with the conventional Macintosh laryngoscope.