Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2008 Feb;34(2):215-21.
doi: 10.1016/j.jcrs.2007.09.022.

Evaluation of iris recognition system for wavefront-guided laser in situ keratomileusis for myopic astigmatism

Affiliations
Comparative Study

Evaluation of iris recognition system for wavefront-guided laser in situ keratomileusis for myopic astigmatism

Sudipta Ghosh et al. J Cataract Refract Surg. 2008 Feb.

Abstract

Purpose: To evaluate the visual and refractive outcomes of wavefront-guided laser in situ keratomileusis (LASIK) using an iris recognition system for the correction of myopic astigmatism.

Setting: Centre for Eye Research Australia, Melbourne Excimer Laser Research Group, and Royal Victorian Eye and Ear Hospital, East Melbourne, Victoria, Australia.

Methods: A comparative analysis of wavefront-guided LASIK was performed with an iris recognition system (iris recognition group) and without iris recognition (control group). The main parameters were uncorrected visual acuity (UCVA), best spectacle-corrected visual acuity, amount of residual cylinder, manifest spherical equivalent (SE), and the index of success using the Alpins method of astigmatism analysis 1 and 3 months postoperatively. A P value less than 0.05 was considered statistically significant.

Results: Preoperatively, the mean SE was -4.32 diopters (D) +/- 1.59 (SD) in the iris recognition group (100 eyes) and -4.55 +/- 1.87 D in the control group (98 eyes) (P = .84). At 3 months, the mean SE was -0.05 +/- 0.21 D and -0.20 +/- 0.40 D, respectively (P = .001), and an SE within +/-0.50 D of emmetropia was achieved in 92.0% and 85.7% of eyes, respectively (P = .07). At 3 months, the UCVA was 20/20 or better in 90.0% and 76.5% of eyes, respectively. A statistically significant difference in the amount of astigmatic correction was seen between the 2 groups (P = .00 and P = .01 at 1 and 3 months, respectively). The index of success was 98.0% in the iris recognition group and 81.6% in the control group (P = .03).

Conclusion: Iris recognition software may achieve better visual and refractive outcomes in wavefront-guided LASIK for myopic astigmatism.