A comparison of radiation exposure with the conventional versus mini C arm in orthopedic extremity surgery
- PMID: 18275738
- DOI: 10.3113/FAI.2008.0058
A comparison of radiation exposure with the conventional versus mini C arm in orthopedic extremity surgery
Abstract
Background: The mini C arm is a mobile fluoroscopic device for use on extremities. It has been reported that this type of device has several advantages over a conventional C arm, the most important of which is a decreased level of radiation scatter to the surgeon.
Methods: This was a case control study. The radiation dose to the surgeon was calculated from the radiation exposure time and distance from the x-ray emitter. Cases using the mini C arm were compared with operation-matched controls that had the identical procedure performed with a conventional C arm.
Results: The mini C arm caused statistically less radiation scatter to the surgeon, but there was no statistically significant difference in the radiation exposure to the patient. This was despite a significantly larger number of exposures with the mini C arm for each operation type.
Conclusions: The mini C arm should be used in preference to the conventional C arm for extremity surgery, if both are available.