Technetium 99m mercaptoacetyltriglycine gamma camera clearance calculations: methodological problems
- PMID: 1828421
- DOI: 10.1007/BF00950751
Technetium 99m mercaptoacetyltriglycine gamma camera clearance calculations: methodological problems
Abstract
Major sources of errors in the gamma-camera methods for the calculation of renal clearance are the accuracy of background correction for obtaining the true renal time-activity curve and the validity of the externally recorded pre-cordial activity as an estimate of the plasmatic time-activity curve. With technetium 99m mercaptoacetyltriglycine (99mTc-MAG3), because of its high protein plasma binding, one could expect minimal extravascular diffusion and hence a more accurate externally detected plasmatic curve. The high extraction rate should reduce the influence of the background, but, on the other hand, the effect of hepatobiliary excretion on the calculation of renal clearance might be significant. Our results suggest that the hepatobiliary excretion of 99mTc-MAG3 does not influence the gamma-camera renal clearance determination, even in patients with low renal function. However, the pre-cordial curve does not reflect accurately the plasmatic disappearance curve; its calibration with a single plasma sample taken at the 20th min is responsible for significant errors, probably because of an unfavourable ratio between the intravascular and extravascular activities at the 20th min.
Similar articles
-
Estimation of technetium-99m-MAG3 renal clearance in children: two gamma camera techniques compared with multiple plasma samples.J Nucl Med. 1991 Sep;32(9):1704-8. J Nucl Med. 1991. PMID: 1831838
-
A comparison of single plasma sample methods to estimate renal clearance using 99mTc-ethylenedicysteine and 99mTc-mercaptoacetyltriglycine.Clin Physiol Funct Imaging. 2013 Sep;33(5):353-8. doi: 10.1111/cpf.12034. Epub 2013 Mar 18. Clin Physiol Funct Imaging. 2013. PMID: 23701132
-
Estimation of technetium-99m-MAG3 plasma clearance in adults from one or two blood samples.J Nucl Med. 1989 Dec;30(12):1955-9. J Nucl Med. 1989. PMID: 2531219
-
The utility of 99mTc-mercaptoacetyltriglycine in captopril renography.Am J Hypertens. 1991 Dec;4(12 Pt 2):731S-736S. doi: 10.1093/ajh/4.12.731s. Am J Hypertens. 1991. PMID: 1837994 Review.
-
Captopril renography. Early observations and diagnostic criteria.Am J Hypertens. 1991 Dec;4(12 Pt 2):678S-684S. doi: 10.1093/ajh/4.12.678s. Am J Hypertens. 1991. PMID: 1837990 Review.
Cited by
-
Quantification of renal haemodynamics with radionuclides.Eur J Nucl Med. 1991;18(4):274-86. doi: 10.1007/BF00186653. Eur J Nucl Med. 1991. PMID: 2070806 Review.
-
Determination of the technetium-99m mercaptoacetyltriglycine plasma clearance in children by means of a single blood sample: a multicentre study. The Paediatric Task Group of the EANM.Eur J Nucl Med. 1993 Mar;20(3):244-8. doi: 10.1007/BF00170006. Eur J Nucl Med. 1993. PMID: 8462614