Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2008 Jul;19(7):725-9.
doi: 10.1111/j.1540-8167.2008.01105.x. Epub 2008 Feb 13.

Validation of a new noninvasive device for the monitoring of peak endocardial acceleration in pigs: implications for optimization of pacing site and configuration

Affiliations

Validation of a new noninvasive device for the monitoring of peak endocardial acceleration in pigs: implications for optimization of pacing site and configuration

Pierre Bordachar et al. J Cardiovasc Electrophysiol. 2008 Jul.

Abstract

Introduction: The peak of endocardial acceleration (PEA) is an index of myocardial contractility. We aimed to (1) demonstrate that the PEA measured by the noninvasive cutaneous precordial application of an accelerometer sensor is related to left ventricular (LV) dP/dt max and (2) assess the usefulness of PEA monitoring during graded ischemia and during different configurations of sequential biventricular pacing.

Methods and results: Measurements of invasive LV dP/dt max were compared with measurements of transcutaneous PEA in seven pigs at baseline and during acute drug infusions; increased heart rate; right, left, biventricular and sequential biventricular pacing before and after graded ischemia induced by the constriction of the left anterior descending coronary artery. A consistent PEA signal was obtained in all animals. PEA changes were highly related to LV dP/dt max changes (r= 0.93; P < 0.001). The changes of LV contractility induced by the different pacing configurations were detected by PEA analysis in the absence of ischemia (r= 0.94; P < 0.001) and in the presence of ischemic LV dysfunction (r= 0.91; P < 0.001).

Conclusion: Noninvasive PEA measurement allows monitoring of left ventricular contractility and may be a useful tool to detect global effect of ventricular ischemia and to optimize the choice of both pacing site and pacing configuration.