Validation of a new noninvasive device for the monitoring of peak endocardial acceleration in pigs: implications for optimization of pacing site and configuration
- PMID: 18284498
- DOI: 10.1111/j.1540-8167.2008.01105.x
Validation of a new noninvasive device for the monitoring of peak endocardial acceleration in pigs: implications for optimization of pacing site and configuration
Abstract
Introduction: The peak of endocardial acceleration (PEA) is an index of myocardial contractility. We aimed to (1) demonstrate that the PEA measured by the noninvasive cutaneous precordial application of an accelerometer sensor is related to left ventricular (LV) dP/dt max and (2) assess the usefulness of PEA monitoring during graded ischemia and during different configurations of sequential biventricular pacing.
Methods and results: Measurements of invasive LV dP/dt max were compared with measurements of transcutaneous PEA in seven pigs at baseline and during acute drug infusions; increased heart rate; right, left, biventricular and sequential biventricular pacing before and after graded ischemia induced by the constriction of the left anterior descending coronary artery. A consistent PEA signal was obtained in all animals. PEA changes were highly related to LV dP/dt max changes (r= 0.93; P < 0.001). The changes of LV contractility induced by the different pacing configurations were detected by PEA analysis in the absence of ischemia (r= 0.94; P < 0.001) and in the presence of ischemic LV dysfunction (r= 0.91; P < 0.001).
Conclusion: Noninvasive PEA measurement allows monitoring of left ventricular contractility and may be a useful tool to detect global effect of ventricular ischemia and to optimize the choice of both pacing site and pacing configuration.
Similar articles
-
Haemodynamic impact of the left ventricular pacing site during graded ischaemia in an open-chest pig model.Europace. 2008 Feb;10(2):242-8. doi: 10.1093/europace/eum285. Europace. 2008. PMID: 18256130
-
Effect of optimizing the VV interval on left ventricular contractility in cardiac resynchronization therapy.Am J Cardiol. 2004 Jun 15;93(12):1500-3. doi: 10.1016/j.amjcard.2004.02.061. Am J Cardiol. 2004. PMID: 15194020
-
Impact of different pacing modes on left ventricular contractility following cardiopulmonary bypass.Pacing Clin Electrophysiol. 2007 Sep;30(9):1083-90. doi: 10.1111/j.1540-8159.2007.00817.x. Pacing Clin Electrophysiol. 2007. PMID: 17725750
-
Does cardiac resynchronisation therapy improve survival and quality of life in patients with end-stage heart failure?Interact Cardiovasc Thorac Surg. 2008 Dec;7(6):1141-6. doi: 10.1510/icvts.2008.183707. Epub 2008 Jun 9. Interact Cardiovasc Thorac Surg. 2008. PMID: 18541605 Review.
-
Noninvasive transcutaneous cardiac pacing.Anesth Analg. 1989 Aug;69(2):229-38. Anesth Analg. 1989. PMID: 2669559 Review. No abstract available.
Cited by
-
Left Ventricular Pressure Estimation Using Machine Learning-Based Heart Sound Classification.Front Cardiovasc Med. 2022 May 25;9:763048. doi: 10.3389/fcvm.2022.763048. eCollection 2022. Front Cardiovasc Med. 2022. PMID: 35694657 Free PMC article.
-
Analysis of Cardiac Vibration Signals Acquired From a Novel Implant Placed on the Gastric Fundus.Front Physiol. 2021 Nov 19;12:748367. doi: 10.3389/fphys.2021.748367. eCollection 2021. Front Physiol. 2021. PMID: 34867453 Free PMC article.
-
Heart failure monitoring with a cardiac resynchronization therapy device-based cardiac contractility sensor: a case series.J Med Case Rep. 2014 Jan 27;8:27. doi: 10.1186/1752-1947-8-27. J Med Case Rep. 2014. PMID: 24467718 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources