A biomechanical and histological analysis of standard versus hydroxyapatite-coated pins for external fixation
- PMID: 18286593
- DOI: 10.1002/jbm.b.31036
A biomechanical and histological analysis of standard versus hydroxyapatite-coated pins for external fixation
Abstract
This sheep study was designed to make a comparative evaluation of two external fixation pin types each with and without hydroxyapatite (HA) coating. The two pins had different taper, pitch, and self drilling capabilities. Twenty Orthofix standard, self-tapping pins (group A), 20 Orthofix HA-coated, self-tapping pins (group B), 20 X-caliber, self-drilling, self-tapping pins (group C), and 20 X-caliber HA-coated, self-drilling, self-tapping pins (group D) were selected. Four pins were implanted in the right femurs of 20 adult sheep that were euthanized at 6 weeks. Mean pin insertion torque was 2745 +/- 822 Nmm in group A, 2726 +/- 784 Nmm in group B, 2818 +/- 552 Nmm in group C, and 2657 +/- 732 Nmm in group D (ns). Mean pin extraction torque was 1567 +/- 541 Nmm in group A, 2524 +/- 838 Nmm in group B, 1650 +/- 650 Nmm in group C, and 2517 +/- 726 Nmm in group D. HA-coated pins (group B and D) had a significantly greater mean pin extraction torque compared to similar uncoated pins (group A and C) (p < 0.0005). Histological analysis showed good osteointegration of the two coated pin types. This study shows that HA-coating is more important for optimal pin fixation than the particular combination of design parameters used in each pin type.
(c) 2007 Wiley Periodicals, Inc.
Similar articles
-
Hydroxyapatite coated external fixation pins: an experimental study.Clin Orthop Relat Res. 1996 Apr;(325):269-75. doi: 10.1097/00003086-199604000-00033. Clin Orthop Relat Res. 1996. PMID: 8998887
-
Fixation strength of tapered versus bicylindrical hydroxyapatite-coated external fixation pins: an animal study.J Biomed Mater Res. 2002;63(1):61-4. doi: 10.1002/jbm.10088. J Biomed Mater Res. 2002. PMID: 11787030
-
Improvement of the bone-pin interface strength in osteoporotic bone with use of hydroxyapatite-coated tapered external-fixation pins. A prospective, randomized clinical study of wrist fractures.J Bone Joint Surg Am. 2001 May;83(5):717-21. doi: 10.2106/00004623-200105000-00010. J Bone Joint Surg Am. 2001. PMID: 11379741 Clinical Trial.
-
State of the art review: techniques to avoid pin loosening and infection in external fixation.J Orthop Trauma. 2002 Mar;16(3):189-95. doi: 10.1097/00005131-200203000-00009. J Orthop Trauma. 2002. PMID: 11880783 Review.
-
Hydroxyapatite-coated external fixation pins.Expert Rev Med Devices. 2005 Jul;2(4):465-71. doi: 10.1586/17434440.2.4.465. Expert Rev Med Devices. 2005. PMID: 16293085 Review.
Cited by
-
From prevention of pin-tract infection to treatment of osteomyelitis during paediatric external fixation.J Child Orthop. 2016 Dec;10(6):605-612. doi: 10.1007/s11832-016-0787-8. Epub 2016 Nov 15. J Child Orthop. 2016. PMID: 27848193 Free PMC article. Review.
-
The effect of HIV infection on the incidence and severity of circular external fixator pin track sepsis: a retrospective comparative study of 229 patients.Strategies Trauma Limb Reconstr. 2014 Aug;9(2):111-5. doi: 10.1007/s11751-014-0194-y. Epub 2014 Jul 24. Strategies Trauma Limb Reconstr. 2014. PMID: 25056512 Free PMC article.
-
Titanium Kirschner Wires Resist Biofilms Better Than Stainless Steel and Hydroxyapatite-coated Wires: An In Vitro Study.Strategies Trauma Limb Reconstr. 2019 May-Aug;14(2):57-64. doi: 10.5005/jp-journals-10080-1426. Strategies Trauma Limb Reconstr. 2019. PMID: 32742415 Free PMC article.
-
Potential of Titanium Pins Coated with Fibroblast Growth Factor-2-Calcium Phosphate Composite Layers to Reduce the Risk of Impaired Bone-Pin Interface Strength in the External Fixation of Distal Radius Fractures.J Clin Med. 2024 May 22;13(11):3040. doi: 10.3390/jcm13113040. J Clin Med. 2024. PMID: 38892751 Free PMC article.
-
Prevention and management of external fixator pin track sepsis.Strategies Trauma Limb Reconstr. 2012 Aug;7(2):67-72. doi: 10.1007/s11751-012-0139-2. Epub 2012 Jun 23. Strategies Trauma Limb Reconstr. 2012. PMID: 22729940 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Miscellaneous