Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2008 Mar;21(1-2):5-14.
doi: 10.1007/s10334-008-0105-7. Epub 2008 Feb 26.

Parallel imaging in non-bijective, curvilinear magnetic field gradients: a concept study

Affiliations

Parallel imaging in non-bijective, curvilinear magnetic field gradients: a concept study

Juergen Hennig et al. MAGMA. 2008 Mar.

Abstract

Objectives: The paper presents a novel and more generalized concept for spatial encoding by non-unidirectional, non- bijective spatial encoding magnetic fields (SEMs). In combination with parallel local receiver coils these fields allow one to overcome the current limitations of neuronal nerve stimulation. Additionally the geometry of such fields can be adapted to anatomy.

Materials and methods: As an example of such a parallel imaging technique using localized gradients (PatLoc)- system, we present a polar gradient system consisting of 2 x 8 rectangular current loops in octagonal arrangement, which generates a radial magnetic field gradient. By inverting the direction of current in alternating loops, a near sinusoidal field variation in the circumferential direction is produced. Ambiguities in spatial assignment are resolved by use of multiple receiver coils and parallel reconstruction. Simulations demonstrate the potential advantages and limitations of this approach.

Results and conclusions: The exact behaviour of PatLoc fields with respect to peripheral nerve stimulation needs to be tested in practice. Based on geometrical considerations SEMs of radial geometry allow for about three times faster gradient switching compared to conventional head gradient inserts and even more compared to whole body gradients. The strong nonlinear geometry of the fields needs to be considered for practical applications.

PubMed Disclaimer

Figures

Fig. 1
Fig. 1
Octagonal coil arrangement used in the simulations. 2 × 8 planar coils are arranged in an octagonal arrangement around the direction of the main field (z). r and ϕ denote the radial resp. azimuthal directions in polar coordinates. Arrows next to the rungs indicate the current direction
Fig. 2
Fig. 2
A color-coded field plot of magnetic field Bz produced by the coil array in (Fig. 1) with a current of 10 A through each of the six rungs at the center of the planar coils at a radius r = 6 cm. The maximum field is 0.68 mT at a radius r = 5.56cm from the center
Fig. 3
Fig. 3
a Field Bz along r. The dotted line represents a constant gradient with identical maximum and minimum. b Gradient G corresponding to the derivative of Bz alongr. The dotted line represents the mean gradient between the field maximum and minimum
Fig. 4
Fig. 4
a Spectrum Fω of a periodically modulated spin distribution with a peak-to-peak distance of 4 pixels and ∼1 pixel halfwidth measured with a constant mean gradient Gconst at a bandwidth ω = 40,960 Hz. The observed halfwidth of the spectrum is ∼2 pixels. b Spectrum Fω of same spin distribution observed in the radial SEMR shown in Fig. 3b at identical bandwidth. c Spectra from a and b mapped to spatial coordinates
Fig. 5
Fig. 5
a Magnetic field profile of SEMC generated from an octopole arrangement of coils with a current of 10 A flowing in alternating directions through the coil configuration shown in Fig. 1. Arrows indicate the primary direction of the field gradient. The wedge shaped outline shows one of the eight subregions for unambiguous (but curved) spatial encoding. b Gradient of the SEMC at equidistant positions in the dotted ring shown in (Fig. 5a) as a function of the azimuth angle ϕ. Only two of the four gradient lobes (0 < ϕ < 180°) are shown for clarity
Fig. 6
Fig. 6
Basic workflow for PatLoc reconstruction of a spin distribution Iρ(x, y) into its PatLoc image Ixy(x, y). For explanation see text
Fig. 7
Fig. 7
Results of simulations of circular (a, b), spokewheel (c, d) and random (e, f) spin distribution patterns imaged by constant Cartesian gradients (left column) versus PatLoc gradients SEMC and SEMR (right column). Linewidth of the modulated geometrical patterns was roughly 1 pixel at 256 × 256 matrix size
Fig. 8
Fig. 8
Spin distribution taken from a 512 × 512 volunteer image imaged simulated to be imaged with conventional gradients (a) and with PatLoc gradients SEMR and SEMC acquired at 384 × 384 (b) and 256 × 256 (c) matrix size
Fig. 9
Fig. 9
a Colour encoded display of the angle α between the gradients produced by the multipolar SEMC and the interleaved multipolar SEMCI generated from SEMC by rotation by 22.5°. b Superposition of isocontour lines of SEMCI and SEMC
Fig. 10
Fig. 10
Simulation of a phantom image produced by combining the circular and spoke wheel patterns in (Fig. 7a, c) when imaged with conventional gradients and with SEMC and SEMCI. Linewidth of the geometrical patterns was ∼1 pixel at 256 × 256 matrix size

References

    1. Schenck JF (2005). Physical interactions of static magnetic fields with living tissues. Prog Biophys Mol Biol 87: 185–204 - DOI - PubMed
    1. Schenck JF (2000). Safety of strong, static magnetic fields. J Magn Reson Imaging 12(1): 2–19 - DOI - PubMed
    1. Hennig J, Zhong K and Speck O (2007). MR-Encephalography: fast multi-channel monitoring of brain physiology with magnetic resonance. Neuroimage 34(1): 212–219 - DOI - PubMed
    1. Lin FH, Wald LL, Ahlfors SP, Hamalainen MS, Kwong KK and Belliveau JW (2006). Dynamic magnetic resonance inverse imaging of human brain function. Magn Reson Med 56(4): 787–802 - DOI - PubMed
    1. Cho ZH and Yi JH (1991). A novel type of surface gradient coil. J Magn Reson 94: 471–495

Publication types