Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2008 Jun;9(2):139-47.
doi: 10.1007/s10561-008-9064-5. Epub 2008 Mar 5.

Validation of 15 kGy as a radiation sterilisation dose for bone allografts manufactured at the Queensland Bone Bank: application of the VDmax 15 method

Affiliations

Validation of 15 kGy as a radiation sterilisation dose for bone allografts manufactured at the Queensland Bone Bank: application of the VDmax 15 method

Huynh Nguyen et al. Cell Tissue Bank. 2008 Jun.

Abstract

Background: ISO 11137-2006 (ISO 11137-2a 2006) provides a VDmax 15 method for substantiation of 15 kGy as radiation sterilisation dose (RSD) for health care products with a relatively low sample requirement. Moreover, the method is also valid for products in which the bioburden level is less than or equal to 1.5. In the literature, the bioburden level of processed bone allografts is extremely low. Similarly, the Queensland Bone Bank (QBB) usually recovers no viable organisms from processed bone allografts. Because bone allografts are treated as a type of health care product, the aim of this research was to substantiate 15 kGy as a RSD for frozen bone allografts at the QBB using method VDmax 15-ISO 11137-2: 2006 (ISO 11137-2e, Procedure for method VDmax 15 for multiple production batches. Sterilisation of health care products - radiation - part 2: establishing the sterilisation dose, 2006; ISO 11137-2f, Procedure for method VDmax 15 for a single production batch. Sterilisation of health care products - radiation - part 2: establishing the sterilisation dose, 2006).

Materials: 30 femoral heads, 40 milled bone allografts and 40 structural bone allografts manufactured according to QBB standard operating procedures were used.

Method: Estimated bioburdens for each bone allograft group were used to calculate the verification doses. Next, 10 samples per group were irradiated at the verification dose, sterility was tested and the number of positive tests of sterility recorded. If the number of positive samples was no more than 1, from the 10 tests carried out in each group, the verification was accepted and 15 kGy was substantiated as RSD for those bone allografts.

Results: The bioburdens in all three groups were 0, and therefore the verification doses were 0 kGy. Sterility tests of femoral heads and milled bones were all negative (no contamination), and there was one positive test of sterility in the structural bone allograft. Accordingly, the verification was accepted.

Conclusion: Using the ISO validated protocol, VDmax 15, 15 kGy was substantiated as RSD for frozen bone allografts manufactured at the QBB.

PubMed Disclaimer

Similar articles

Cited by

Publication types

LinkOut - more resources