A novel approach for quantitative analysis of intracoronary optical coherence tomography: high inter-observer agreement with computer-assisted contour detection
- PMID: 18324698
- DOI: 10.1002/ccd.21482
A novel approach for quantitative analysis of intracoronary optical coherence tomography: high inter-observer agreement with computer-assisted contour detection
Abstract
Objective: This study aims to examine observer-related variability of quantitative optical coherence tomography (OCT) derived measurements from both in vitro and in vivo pullback data.
Background: Intravascular OCT is a new imaging modality using infrared light and offering 10 times higher image resolution (15 microm) compared to intravascular ultrasound. The quantitative analysis of in vivo intracoronary OCT imaging is complicated by the presence of blood, motion artifacts and the large quantity of information that has to be processed.
Methods: We developed a standardized, automated quantification process for intracoronary OCT pullback data with inter-observer variability assessed both in vitro by using postmortem human coronary arteries and in vivo by studying simple and complex coronary pathology and outcomes following stent implantation. The consensus between measurements by two observers was analyzed using the intraclass and interclass correlation coefficient and the reliability coefficients. Bland-Altman plots were generated to assess the relationship between variability and absolute measurements.
Results: In vitro OCT assessment was performed in nine postmortem coronary arteries. The time needed for semiautomated contour detection of a 15-mm long coronary segment was approximately 40 min. The absolute and relative difference between lumen area measurements derived from two observers was low [0.02 +/- 0.10 mm(2); (0.3 +/- 0.5)% respectively] with excellent correlation confirmed by linear regression analysis (R(2) = 0.99; P < 0.001). Similarly, in vivo measurements demonstrated a high correlation with the main source of inter-observer variation occurring as a result of coronary dissection and motion artifact. The absolute and relative difference between measurements were 0.11 +/- 0.33 mm(2) (1.57 +/- 0.05)% for lumen area (R(2) = 0.98; P < 0.001), 0.17 +/- 0.68 mm(2) (1.44 +/- 0.08)% for stent area (R(2) = 0.94; P < 0.001), and 0.26 +/- 0.72 mm(2) (14.08 +/- 0.37)% for neointimal area (R(2) = 0.78; P < 0.001).
Conclusions: Highly accurate computer-assisted quantitative analysis ofintracoronary OCT pullbacks is feasible with low inter-observer variability. The presented approach allows for observer independent analysis of detailed vessel structures, and may be a valuable tool for future longitudinal studies incorporating OCT.
(c) 2008 Wiley-Liss, Inc.
Comment in
-
What can we expect from quantitative OCT?Catheter Cardiovasc Interv. 2008 Aug 1;72(2):236. doi: 10.1002/ccd.21704. Catheter Cardiovasc Interv. 2008. PMID: 18655154 No abstract available.
Similar articles
-
Fully automatic three-dimensional quantitative analysis of intracoronary optical coherence tomography: method and Validation.Catheter Cardiovasc Interv. 2009 Dec 1;74(7):1058-65. doi: 10.1002/ccd.22125. Catheter Cardiovasc Interv. 2009. PMID: 19521990
-
Accuracy and reproducibility of stent-strut thickness determined by optical coherence tomography.J Invasive Cardiol. 2009 Nov;21(11):602-5. J Invasive Cardiol. 2009. PMID: 19901417
-
Assessment of culprit lesion morphology in acute myocardial infarction: ability of optical coherence tomography compared with intravascular ultrasound and coronary angioscopy.J Am Coll Cardiol. 2007 Sep 4;50(10):933-9. doi: 10.1016/j.jacc.2007.04.082. Epub 2007 Aug 20. J Am Coll Cardiol. 2007. PMID: 17765119
-
Optical coherence tomography: high resolution intravascular imaging to evaluate vascular healing after coronary stenting.Catheter Cardiovasc Interv. 2008 Aug 1;72(2):237-47. doi: 10.1002/ccd.21606. Catheter Cardiovasc Interv. 2008. PMID: 18655155 Review.
-
Optical coherence tomography.Cardiovasc Radiat Med. 2003 Oct-Dec;4(4):198-204. doi: 10.1016/j.carrad.2003.12.003. Cardiovasc Radiat Med. 2003. PMID: 15321058 Review.
Cited by
-
Optimized Computer-Aided Segmentation and Three-Dimensional Reconstruction Using Intracoronary Optical Coherence Tomography.IEEE J Biomed Health Inform. 2018 Jul;22(4):1168-1176. doi: 10.1109/JBHI.2017.2762520. IEEE J Biomed Health Inform. 2018. PMID: 29969405 Free PMC article.
-
Interstudy reproducibility of the second generation, Fourier domain optical coherence tomography in patients with coronary artery disease and comparison with intravascular ultrasound: a study applying automated contour detection.Int J Cardiovasc Imaging. 2013 Jan;29(1):39-51. doi: 10.1007/s10554-012-0067-8. Epub 2012 May 26. Int J Cardiovasc Imaging. 2013. PMID: 22639296 Free PMC article.
-
Optical coherence tomography provides images similar to histology and allows the performance of extensive measurements of drug-eluting metal stents in animal ureters.Lasers Med Sci. 2014 Jul;29(4):1453-62. doi: 10.1007/s10103-014-1548-x. Epub 2014 Mar 4. Lasers Med Sci. 2014. PMID: 24590241
-
Optical Coherence Tomography: Potential Clinical Applications.Curr Cardiovasc Imaging Rep. 2012 Aug;5(4):206-220. doi: 10.1007/s12410-012-9140-x. Epub 2012 May 3. Curr Cardiovasc Imaging Rep. 2012. PMID: 22798978 Free PMC article.
-
The diagnostic value of intracoronary optical coherence tomography.Herz. 2011 Aug;36(5):417-29. doi: 10.1007/s00059-011-3487-7. Herz. 2011. PMID: 21744151 Review.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources