Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2008 Apr;20(4):342-8.
doi: 10.1097/MEG.0b013e3282f2a5cf.

Prospective pilot evaluation of a new needle prototype for endoscopic ultrasonography-guided fine-needle aspiration: comparison of cytology and histology yield

Affiliations

Prospective pilot evaluation of a new needle prototype for endoscopic ultrasonography-guided fine-needle aspiration: comparison of cytology and histology yield

Ioannis S Papanikolaou et al. Eur J Gastroenterol Hepatol. 2008 Apr.

Abstract

Objectives: Endoscopic ultrasonography (EUS) with the adjunct of EUS-guided fine needle aspiration has become an important diagnostic modality in gastroenterologic oncology. EUS-guided fine needle aspiration mainly relies on cytology; data are scarce that compare cytology and histology. While testing a 22-gauge prototype needle, we prospectively compared the yield for both.

Methods: Forty-two consecutive patients (27 male, 15 female; mean age 59.2 years, range: 17-90 years) were included. In each patient we aimed to make two needle passes, and if the material acquired appeared insufficient macroscopically (no in-room cytopathology was available), further passes were done. The material was sent for cytological and histological assessment.

Results: A median number of two passes (range: 2-3) were uneventfully performed for pancreatic lesions (n=30), mediastinal and other lymph nodes/masses (n=8) and various other lesions (n=4) and yielded adequate material for cytology, histology or at least one of the two investigations in 62, 67 and 74% of patients, respectively. No false positive results were found (specificity 100%). Sensitivities were 58.6 and 65.5%, respectively, for cytology and histology alone; combined assessment increased sensitivity to 79.3%. When adjusted values were calculated, based only on those cases with adequate material, sensitivity was 89.5% for cytology and 85.7% for histology, and increased to 100% with combined assessment.

Conclusion: The new needle achieves sensitivities similar to those previously reported with no significant differences in sensitivity between cytology and histology. More effective tissue acquisition methods must be sought to improve overall results.