A fluidics comparison of Alcon Infiniti, Bausch & Lomb Stellaris, and Advanced Medical Optics Signature phacoemulsification machines
- PMID: 18343350
- DOI: 10.1016/j.ajo.2008.01.024
A fluidics comparison of Alcon Infiniti, Bausch & Lomb Stellaris, and Advanced Medical Optics Signature phacoemulsification machines
Abstract
Purpose: To compare three phacoemulsification machines for measurement accuracy and postocclusion surge (POS) in human cadaver eyes.
Design: In vitro comparisons of machine accuracy and POS.
Methods: Tip vacuum and flow were compared with machine indicated vacuum and flow. All machines were placed in two human cadaver eyes and POS was determined.
Results: Vacuum (% of actual) was 101.9% +/- 1.7% for Infiniti (Alcon, Fort Worth, Texas, USA), 93.2% +/- 3.9% for Stellaris (Bausch & Lomb, Rochester, New York, USA), and 107.8% +/- 4.6% for Signature (Advanced Medical Optics, Santa, Ana, California, USA; P < .0001). At 60 ml/minute flow, actual flow and unoccluded flow vacuum (UFV) was 55.8 +/- 0.4 ml/minute and 197.7 +/- 0.7 mm Hg for Infiniti, 53.5 +/- 0.0 ml/minute and 179.8 +/- 0.9 mm Hg for Stellaris, and 58.5 +/- 0.0 ml/minute and 115.1 +/- 2.3 mm Hg for Signature (P < .0001). POS in an 32-year-old eye was 0.33 +/- 0.05 mm for Infiniti, 0.16 +/- 0.06 mm for Stellaris, and 0.13 +/- 0.04 mm for Signature at 550 mm Hg, 60 cm bottle height, 45 ml/minute flow with 19-gauge tips (P < .0001 for Infiniti vs Stellaris and Signature). POS in an 81-year-old eye was 1.51 +/- 0.22 mm for Infiniti, 0.83 +/- 0.06 mm for Stellaris, 0.67 +/- 0.01 mm for Signature at 400 mm Hg vacuum, 70 cm bottle height, 40 ml/minute flow with 19-gauge tips (P < .0001).
Conclusions: Machine-indicated accuracy, POS, and UFV were statistically significantly different. Signature had the lowest POS and vacuum to maintain flow. Regarding POS, Stellaris was close to Signature; regarding vacuum to maintain flow, Infiniti and Stellaris were similar. Minimizing POS and vacuum to maintain flow potentially are important in avoiding ocular damage and surgical complications.
Similar articles
-
Objective measurement of postocclusion surge during phacoemulsification in human eye-bank eyes.Am J Ophthalmol. 2007 Mar;143(3):437-40. doi: 10.1016/j.ajo.2006.11.017. Epub 2006 Dec 18. Am J Ophthalmol. 2007. PMID: 17222793
-
Phacodynamics: an aspiration flow vs vacuum comparison.Am J Ophthalmol. 2006 Aug;142(2):320-2. doi: 10.1016/j.ajo.2006.02.034. Am J Ophthalmol. 2006. PMID: 16876517
-
Fluidics and heat generation of Alcon Infiniti and Legacy, Bausch & Lomb Millennium, and advanced medical optics sovereign phacoemulsification systems.Am J Ophthalmol. 2006 Sep;142(3):387-92. doi: 10.1016/j.ajo.2006.04.042. Am J Ophthalmol. 2006. PMID: 16935581
-
Machines and Cutters: Stellaris PC.Dev Ophthalmol. 2014;54:8-16. doi: 10.1159/000360443. Epub 2014 Aug 26. Dev Ophthalmol. 2014. PMID: 25196746 Review.
-
Phaco fluidics and phaco ultrasound power modulations.Ophthalmol Clin North Am. 2006 Dec;19(4):457-68. doi: 10.1016/j.ohc.2006.08.001. Ophthalmol Clin North Am. 2006. PMID: 17067901 Review.
Cited by
-
Comparison of occlusion break responses and vacuum rise times of phacoemulsification systems.BMC Ophthalmol. 2014 Jul 30;14:96. doi: 10.1186/1471-2415-14-96. BMC Ophthalmol. 2014. PMID: 25074069 Free PMC article.
-
Prospective randomized comparative study between venturi and peristaltic pumps in WhiteStar Signature® phacoemulsification machine.Clin Ophthalmol. 2018 Dec 27;13:49-52. doi: 10.2147/OPTH.S177978. eCollection 2019. Clin Ophthalmol. 2018. PMID: 30643379 Free PMC article.
-
Phaco-efficiency test and re-aspiration analysis of repulsed particle in phacoemulsification.Graefes Arch Clin Exp Ophthalmol. 2013 Apr;251(4):1157-61. doi: 10.1007/s00417-012-2240-0. Epub 2013 Jan 3. Graefes Arch Clin Exp Ophthalmol. 2013. PMID: 23280561
-
Efficacy and safety of femtosecond laser-assisted cataract surgery versus conventional phacoemulsification for cataract: a meta-analysis of randomized controlled trials.Sci Rep. 2015 Aug 13;5:13123. doi: 10.1038/srep13123. Sci Rep. 2015. PMID: 26269445 Free PMC article.
-
Comparison of surgical parameters using different lens fragmentation patterns in eyes undergoing laser-assisted cataract surgery.Int Ophthalmol. 2019 Nov;39(11):2459-2465. doi: 10.1007/s10792-019-01087-8. Epub 2019 Mar 4. Int Ophthalmol. 2019. PMID: 30830545 Clinical Trial.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical