. 2008 Mar 24;168(6):663-6.
doi: 10.1001/archinternmed.2007.131.

Influence of local reference populations on upper limits of normal for serum alanine aminotransferase levels

Collaborators

Influence of local reference populations on upper limits of normal for serum alanine aminotransferase levels

Brent A Neuschwander-Tetri et al. Arch Intern Med. .
No abstract available

PubMed Disclaimer

Conflict of interest statement

The authors report no financial conflicts of interest.

Figures

Figure 1
Figure 1
Reference ranges for serum ALT reported by the 11 NASH CRN laboratories. The bars indicate the reference ranges; the upper limit for women varied from 31 U/L in laboratory B to 55 U/L in laboratory E; the upper limit for men varied from 35 U/L in laboratory A to 79 U/L in laboratory K. The lower limit ranged from 0 to 12 U/L.
Figure 2
Figure 2
Mean ALT values for CAP sample C15. The gray bars indicate the laboratory specific means of replicate analyses and the thin line error bars indicate ± 1 SD in each direction; four laboratories (A, D, J, K) did not provide standard deviations of their results. The national machine specific mean ± 1 SD in each direction for the analyzer used in each laboratory is indicated by overlying the thick line error bars. The results from most laboratories were within 2 U/L of the national mean for their specific analyzer; laboratory A deviated the most from the national mean yet the difference was still only 4.9 U/L, indicating that laboratory performance was generally excellent.
Figure 3
Figure 3
The ALT activity of one standardized sample (C15) reported by 11 different laboratories is expressed as a multiple of the gender-specific ULN for those laboratories. The disparate ULN values reported by different laboratories causes the results of this one sample to be variably interpreted as 0.66 to 1.75 × ULN if it came from a man and 0.87 to 1.73 × ULN if it came from a woman. If this sample was from a woman, it would be reported as normal by 3 laboratories (E, F, G) and elevated by the other 8 (Figure 3). If it was from a man, it would be reported as normal in 7 laboratories and elevated in 4 (A, B, C, D)

References

    1. Canalias F, Camprubi S, Sanchez M, Gella FJ. Metrological traceability of values for catalytic concentration of enzymes assigned to a calibration material. Clin Chem Lab Med. 2006;44:333–339. - PubMed
    1. PetitClerc C, Wilding P. The theory of reference values. Part 2. Selection of individuals for the production of reference values. J Clin Chem Clin Biochem. 1984;22:203–208. - PubMed
    1. Prati D, Taioli E, Zanella A, et al. Updated definitions of healthy ranges for serum alanine aminotransferase levels. Ann Intern Med. 2002;137:1–9. - PubMed
    1. Kaplan MM. Alanine aminotransferase levels: what’s normal? Ann Intern Med. 2002;137:49–51. - PubMed
    1. Kunde SS, Lazenby AJ, Clements RH, Abrams GA. Spectrum of NAFLD and diagnostic implications of the proposed new normal range for serum ALT in obese women. Hepatology. 2005;42:650–656. - PubMed