Bioresorbable polylactide interbody implants in an ovine anterior cervical discectomy and fusion model: three-year results
- PMID: 18379399
- DOI: 10.1097/BRS.0b013e3181695716
Bioresorbable polylactide interbody implants in an ovine anterior cervical discectomy and fusion model: three-year results
Abstract
Study design: In vivo study of anterior discectomy and fusion using a bioresorbable 70:30 poly(l-lactide-co-d,l-lactide) interbody implant in an ovine model.
Objective: To evaluate the efficacy of the polylactide implant to function as an interbody fusion device, and to assess the tissue reaction to the material during the resorption process.
Summary of background data: The use of polylactide as a cervical interbody implant has several potential advantages when compared with traditional materials. Having an elastic modulus very similar to bone minimizes the potential for stress shielding, and as the material resorbs additional loading is transferred to the developing fusion mass. Although preclinical and clinical studies have demonstrated the suitability of polylactide implants for lumbar interbody fusion, detailed information on cervical anterior cervical discectomy and fusion (ACDF) with polylactide devices is desirable.
Methods: Single level ACDF was performed in 8 skeletally mature ewes. Bioresorbable 70:30 poly (l-lactide-co-d,l-lactide) interbody implants packed with autograft were used with single-level metallic plates. Radiographs were made every 3 months up to 1 year, and yearly thereafter. The animals were killed at 6 months (3 animals), 12 months (3 animals), and 36 months (2 animals). In addition to the serial plain radiographs, the specimens were evaluated by nondestructive biomechanical testing and undecalcified histologic analysis.
Results: The bioresorbable polylactide implants were effective in achieving interbody fusion. The 6-month animals appeared fused radiographically and biomechanically, whereas histologic sections demonstrated partial fusion (in 3 of 3 animals). Radiographic fusion was confirmed histologically and biomechanically at 12 months (3 of 3 animals) and 36 months (2 of 2 animals). A mild chronic inflammatory response to the resorbing polylactide implant was observed at both 6 months and 12 months. At 36 months, the operative levels were solidly fused and the implants were completely resorbed. No adverse tissue response was observed in any animal at any time period.
Conclusion: Interbody fusion was achieved using bioresorbable polylactide implants, with no evidence of implant collapse, extrusion, or adverse tissue response to the material. The use of polylactide as a cervical interbody device appears both safe and effective based on these ACDF animal model results.
Similar articles
-
Failure of resorbable plates and screws in an ovine model of anterior cervical discectomy and fusion.Spine J. 2011 Sep;11(9):876-83. doi: 10.1016/j.spinee.2011.06.016. Epub 2011 Jul 20. Spine J. 2011. PMID: 21775217
-
Bioabsorbable interbody cages in a sheep cervical spine fusion model.Spine (Phila Pa 1976). 2004 Sep 1;29(17):1845-55; discussion 1856. doi: 10.1097/01.brs.0000137060.79732.78. Spine (Phila Pa 1976). 2004. PMID: 15534403
-
Radiographic, histologic, and chemical evaluation of bioresorbable 70/30 poly-L-lactide-CO-D, L-lactide interbody fusion cages in a goat model.Spine (Phila Pa 1976). 2006 Jun 15;31(14):1559-67. doi: 10.1097/01.brs.0000221984.12004.3b. Spine (Phila Pa 1976). 2006. PMID: 16778688
-
Anterior cervical discectomy and fusion using a porous hydroxyapatite bone graft substitute.Spine (Phila Pa 1976). 1994 Oct 15;19(20):2348-57. doi: 10.1097/00007632-199410150-00017. Spine (Phila Pa 1976). 1994. PMID: 7531370 Review.
-
Anterior cervical discectomy and fusion: review and update for radiologists.Skeletal Radiol. 2018 Jan;47(1):7-17. doi: 10.1007/s00256-017-2798-z. Epub 2017 Oct 23. Skeletal Radiol. 2018. PMID: 29058045 Review.
Cited by
-
Biphasic calcium phosphate contained within a polyetheretherketone cage with and without plating for anterior cervical discectomy and fusion.Orthop Surg. 2012 Aug;4(3):156-65. doi: 10.1111/j.1757-7861.2012.00185.x. Orthop Surg. 2012. PMID: 22927149 Free PMC article.
-
The use of BoneWelding® technology in spinal surgery: an experimental study in sheep.Eur Spine J. 2011 Nov;20(11):1821-36. doi: 10.1007/s00586-011-1799-1. Epub 2011 Apr 27. Eur Spine J. 2011. PMID: 21523457 Free PMC article.
-
Static and dynamic fatigue behavior of topology designed and conventional 3D printed bioresorbable PCL cervical interbody fusion devices.J Mech Behav Biomed Mater. 2015 Sep;49:332-42. doi: 10.1016/j.jmbbm.2015.05.015. Epub 2015 May 27. J Mech Behav Biomed Mater. 2015. PMID: 26072198 Free PMC article.
-
In Vivo and In Vitro Study of a Polylactide-Fiber-Reinforced β-Tricalcium Phosphate Composite Cage in an Ovine Anterior Cervical Intercorporal Fusion Model.Int J Biomater. 2011;2011:109638. doi: 10.1155/2011/109638. Epub 2011 Oct 26. Int J Biomater. 2011. PMID: 22114603 Free PMC article.
-
Bioabsorbable self-retaining PLA/nano-sized β-TCP cervical spine interbody fusion cage in goat models: an in vivo study.Int J Nanomedicine. 2017 Oct 3;12:7197-7205. doi: 10.2147/IJN.S132041. eCollection 2017. Int J Nanomedicine. 2017. PMID: 29042769 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Research Materials