Measured attenuation correction using the Maximum Likelihood algorithm
- PMID: 1839850
Measured attenuation correction using the Maximum Likelihood algorithm
Abstract
Quantitative determination of local radioactivity concentration in positron emission tomography (PET) requires a good attenuation correction procedure to reconstruct the emission image. Using a similar Maximum Likelihood (ML) algorithm as for the reconstruction of the emission image, a method is proposed to reconstruct a transmission image, i.e. a map of absorption coefficients. This reconstructed transmission image is then used to calculate the attenuation correction factors needed for the ML reconstruction of the emission image. This approach takes automatically into account the convolution step in the acquisition process (caused by various smoothing factors, e.g. the detector width). This results in appreciable noise suppression without loss of resolution due to filtering, thus making the reconstructed images easier to interpret. A comparison is made with other estimates for the measured attenuation correction using phantom studies.
Similar articles
-
Development and evaluation of MRI based Bayesian image reconstruction methods for PET.Comput Med Imaging Graph. 2004 Jun;28(4):177-84. doi: 10.1016/j.compmedimag.2003.11.005. Comput Med Imaging Graph. 2004. PMID: 15121207
-
[A new Bayesian reconstruction algorithm for PET images based on correction of the detected sinogram data].Nan Fang Yi Ke Da Xue Xue Bao. 2007 Mar;27(3):325-8. Nan Fang Yi Ke Da Xue Xue Bao. 2007. PMID: 17425984 Chinese.
-
NEMA NU4-2008 image quality performance report for the microPET focus 120 and for various transmission and reconstruction methods.J Nucl Med. 2009 Oct;50(10):1730-8. doi: 10.2967/jnumed.109.063974. Epub 2009 Sep 16. J Nucl Med. 2009. PMID: 19759103
-
Determination of the attenuation map in emission tomography.J Nucl Med. 2003 Feb;44(2):291-315. J Nucl Med. 2003. PMID: 12571222 Review.
-
X-ray-based attenuation correction for positron emission tomography/computed tomography scanners.Semin Nucl Med. 2003 Jul;33(3):166-79. doi: 10.1053/snuc.2003.127307. Semin Nucl Med. 2003. PMID: 12931319 Review.