Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2008 Apr;64(4):990-5.
doi: 10.1097/TA.0b013e31802bf051.

The effect of pilot hole size on the insertion torque and pullout strength of self-tapping cortical bone screws in osteoporotic bone

Affiliations

The effect of pilot hole size on the insertion torque and pullout strength of self-tapping cortical bone screws in osteoporotic bone

Suneel Battula et al. J Trauma. 2008 Apr.

Abstract

Background: All surgical screws can experience failure if the torsional, tensile, and flexion loads exerted on the screws are excessively high. The use of self-tapping screws (STS) results in higher insertion torques (IT) as these screws cut their own threads in the pilot hole drilled in the bone. In this study, the torque for inserting the STS into an osteoporotic bone block for different pilot hole sizes (PHS) was measured and the pullout strength (PS) for extraction of the screws was determined for different depths of insertion, 0 mm, 1 mm, and 2 mm beyond the far cortex.

Methods: Seventy-two Synthes stainless steel STS (40 mm length and 3.5 mm diameter) were inserted into pilot holes of sizes 2.55 (A: 73% OD), 2.50 (B: 71.5%), 2.45 (C: 70%), and 2.8 mm (D: 80%). Using a digital torque screwdriver, screws were inserted to 0 mm, 1 mm or 2 mm past the far cortex. Pullout tests were conducted with an Instron materials testing system. Analysis of variance and Student-Neuman-Keuls tests were performed to determine the effect of DOI and PHS on the loading energy, PS, and IT.

Results: Results demonstrated that IT of the screws inserted into pilot holes A, B, and C were higher than those in D. It was also observed that PS and loading energy for 1 mm and 2 mm penetration past the far cortex were higher than those for 0 mm regardless of PHS. This study also found that an increase in PHS to 2.8 mm will reduce IT but will also reduce the PS relative to a PHS of 2.5 mm, the current standard for 3.5 mm screws.

Conclusions: The results of previously published studies regarding the effect of pilot hole size on PS in healthy cortical bone cannot be applied to the osteoporotic environment. The findings presented in this research support using PHS no larger than 71.5% of the screw outer diameter (i.e., pilot hole size of 2.5 mm for 3.5 mm screws) and inserting screws at least 2 mm beyond the far cortex to maximize PS and minimize iatrogenic damage in osteoporotic bone.

PubMed Disclaimer

Substances