A Bayesian precision profile for measuring the quality of immunoassay experiments
- PMID: 18407899
- DOI: 10.1098/rsta.2008.0034
A Bayesian precision profile for measuring the quality of immunoassay experiments
Abstract
A precision profile, the relationship between the concentration of a substance and its measured precision, is a convenient way of conveying the ability of an immunoassay to accurately measure the concentration of a substance in blood serum. A precision profile is characterized by the definition of precision. Historically, precision has been evaluated as the standard error of an estimator of the concentration in a sample conditional on the true concentration. In this paper, Bayesian predictive inference is used to develop a new measure of precision based on the accuracy with which an assay could infer the concentration in a hypothetical new sample. This leads to a natural procedure for evaluating a precision profile that avoids using approximations such as those inherent in traditional methods.
Similar articles
-
The effect of serial dilution error on calibration inference in immunoassay.Biometrics. 1998 Mar;54(1):19-32. Biometrics. 1998. PMID: 9544505
-
The effect of variance function estimation on nonlinear calibration inference in immunoassay data.Biometrics. 1996 Mar;52(1):158-75. Biometrics. 1996. PMID: 8934590
-
Can chemiluminescent immunoanalysis of thyroid hormones stand for a reference method?Sb Lek. 1994;95(4):363-72. Sb Lek. 1994. PMID: 8867708
-
Thyroglobulin measurement. Techniques, clinical benefits, and pitfalls.Endocrinol Metab Clin North Am. 1995 Dec;24(4):841-63. Endocrinol Metab Clin North Am. 1995. PMID: 8608781 Review.
-
Apolipoprotein assays: standardization and quality control.Scand J Clin Lab Invest Suppl. 1990;198:58-65. Scand J Clin Lab Invest Suppl. 1990. PMID: 2111937 Review.
Cited by
-
BPA Endocrine Disruptor Detection at the Cutting Edge: FPIA and ELISA Immunoassays.Biosensors (Basel). 2023 Jun 19;13(6):664. doi: 10.3390/bios13060664. Biosensors (Basel). 2023. PMID: 37367029 Free PMC article.
-
Fluorescence polarization immunoassay for the determination of diclofenac in wastewater.Anal Bioanal Chem. 2021 Feb;413(4):999-1007. doi: 10.1007/s00216-020-03058-w. Epub 2020 Nov 25. Anal Bioanal Chem. 2021. PMID: 33241445 Free PMC article.
-
Sophisticated diagnostic modalities.Philos Trans A Math Phys Eng Sci. 2008 Jul 13;366(1874):2251-2. doi: 10.1098/rsta.2008.0046. Philos Trans A Math Phys Eng Sci. 2008. PMID: 18407891 Free PMC article. No abstract available.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources