Reducing radiation dose in rest-stress cardiac PET/CT by single poststress cine CT for attenuation correction: quantitative validation
- PMID: 18413384
- DOI: 10.2967/jnumed.107.049163
Reducing radiation dose in rest-stress cardiac PET/CT by single poststress cine CT for attenuation correction: quantitative validation
Abstract
Cardiac PET/CT is optimized by cine CT with dedicated shift software for manual correction of attenuation-emission misregistration. Separate rest and stress CT scans incur greater radiation dose to patients than does standard helical PET/CT or "pure" PET using rotating rod attenuation sources. To reduce radiation dose, we tested quantitative accuracy of using a single poststress cine CT attenuation scan for reconstructing rest perfusion images to eliminate resting CT attenuation scans.
Methods: A total of 250 consecutive patients underwent diagnostic rest-dipyridamole myocardial perfusion PET/CT with (82)Rb and a 16-slice PET/CT scanner using averaged cine CT attenuation data during breathing at rest and stress. After correcting for any attenuation-emission misregistration, we quantitatively compared resting perfusion images reconstructed using rest cine CT attenuation data with the same resting emission data reconstructed with poststress cine CT attenuation data. Automated software quantifying average regional quadrant activity, severity, size, and combined size and severity of perfusion defects was used for this comparison.
Results: Resting perfusion images reconstructed using rest cine CT attenuation data were quantitatively comparable to resting images reconstructed with poststress cine CT attenuation data with no clinically significant differences. Twenty-five (10%) of 250 cases required shifting of stress cine CT attenuation data to achieve optimal attenuation-emission coregistration with resting perfusion data. Eliminating rest CT attenuation scans reduced CT radiation dose by 50% below rest-plus-stress cine CT protocols.
Conclusion: Resting perfusion images reconstructed using poststress cine CT attenuation data are quantitatively comparable to resting images reconstructed with resting cine CT attenuation data. Eliminating the rest CT scan reduces CT radiation dose by 50%.
Similar articles
-
Frequent diagnostic errors in cardiac PET/CT due to misregistration of CT attenuation and emission PET images: a definitive analysis of causes, consequences, and corrections.J Nucl Med. 2007 Jul;48(7):1112-21. doi: 10.2967/jnumed.107.039792. Epub 2007 Jun 15. J Nucl Med. 2007. PMID: 17574974
-
Common artifacts in PET myocardial perfusion images due to attenuation-emission misregistration: clinical significance, causes, and solutions.J Nucl Med. 2004 Jun;45(6):1029-39. J Nucl Med. 2004. PMID: 15181138 Clinical Trial.
-
Coronary calcium score scan-based attenuation correction in cardiovascular PET imaging.Nucl Med Commun. 2010 Sep;31(9):780-7. doi: 10.1097/MNM.0b013e32833c3117. Nucl Med Commun. 2010. PMID: 20559155
-
X-ray-based attenuation correction for positron emission tomography/computed tomography scanners.Semin Nucl Med. 2003 Jul;33(3):166-79. doi: 10.1053/snuc.2003.127307. Semin Nucl Med. 2003. PMID: 12931319 Review.
-
New cardiac cameras: single-photon emission CT and PET.Semin Nucl Med. 2014 Jul;44(4):232-51. doi: 10.1053/j.semnuclmed.2014.04.003. Semin Nucl Med. 2014. PMID: 24948149 Review.
Cited by
-
Average CT in PET studies of colorectal cancer patients with metastasis in the liver and esophageal cancer patients.J Appl Clin Med Phys. 2010 Feb 4;11(1):3073. doi: 10.1120/jacmp.v11i1.3073. J Appl Clin Med Phys. 2010. PMID: 20160693 Free PMC article.
-
Dose calculation with respiration-averaged CT processed from cine CT without a respiratory surrogate.Med Phys. 2008 Dec;35(12):5738-47. doi: 10.1118/1.3015197. Med Phys. 2008. PMID: 19175131 Free PMC article.
-
Diagnosing coronary artery disease with hybrid PET/CT: it takes two to tango.J Nucl Cardiol. 2013 Oct;20(5):874-90. doi: 10.1007/s12350-013-9753-8. J Nucl Cardiol. 2013. PMID: 23842709 Review.
-
Computed tomography calcium score scan for attenuation correction of N-13 ammonia cardiac positron emission tomography: effect of respiratory phase and registration method.Int J Cardiovasc Imaging. 2013 Aug;29(6):1351-60. doi: 10.1007/s10554-013-0207-9. Epub 2013 Mar 17. Int J Cardiovasc Imaging. 2013. PMID: 23504215
-
Partial volume correction incorporating Rb-82 positron range for quantitative myocardial perfusion PET based on systolic-diastolic activity ratios and phantom measurements.J Nucl Cardiol. 2011 Apr;18(2):247-58. doi: 10.1007/s12350-010-9327-y. Epub 2010 Dec 24. J Nucl Cardiol. 2011. PMID: 21184208 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical