How to improve performance of robotic totally endoscopic coronary artery bypass grafting
- PMID: 18424293
- DOI: 10.1016/j.amjsurg.2007.11.010
How to improve performance of robotic totally endoscopic coronary artery bypass grafting
Abstract
Background: We investigated whether specific surgical measures during the learning curve can influence procedural performance in robotic totally endoscopic coronary artery bypass grafting (TECAB).
Methods: From 2001 to 2006, 70 patients underwent TECAB using the da Vinci telemanipulation system (Intuitive Surgical, Sunnyvale, CA). The following measures were employed with the intent to improve procedure performance: (1) introduction of a fixed team of surgeons at case 14; (2) application of fibrin glue to seal the anastomosis at case 28; and (3) use of a fourth port for transthoracic assistance at case 49.
Results: All 3 measures resulted in a reduction of operating room (OR) time (27%, 20%, and 20%). Use of a stable OR team and peri-anastomotic fibrin glue reduced the rate of conversions and on-table revisions from 39% to 9% (P = .006) and from 26% to 7% (P = .038), respectively. Peri-anastomotic fibrin glue significantly reduced postoperative revisions for bleeding. Using transthoracic assistance, anastomotic time was shortened from 35 minutes (range 23-67) to 25 minutes (range 16-100) (P < .001).
Conclusion: The 3 measures can improve intraoperative and clinical results in TECAB.