. 2005 Jun:Chapter 8:Unit 8.8.
doi: 10.1002/0471140864.ps0808s40.

Expanded-bed adsorption chromatography

Affiliations

Expanded-bed adsorption chromatography

Robert M Kennedy. Curr Protoc Protein Sci. 2005 Jun.

Abstract

Expanded-bed adsorption (EBA) chromatography is a convenient and effective technique for the capture of proteins directly from unclarified crude sample. In EBA chromatography, the settled bed is first expanded by upward flow of equilibration buffer. The crude feed, a mixture of soluble proteins, contaminants, cells, and cell debris, is then passed upward through the expanded bed. Target proteins are captured on the adsorbent, while particulates and contaminants pass through. A change to elution buffer while maintaining upward flow results in desorption of the target protein in expanded-bed mode. Alternatively, if the flow is reversed, the adsorbed particles will quickly settle and the proteins can be desorbed by an elution buffer. The mode used for elution (expanded-bed versus settled-bed) depends on the characteristics of the feed. After elution, the adsorbent is cleaned with a predefined cleaning-in-place (CIP) solution, with cleaning followed by either column regeneration (for further use) or storage.

PubMed Disclaimer

MeSH terms

LinkOut - more resources