How accurately, reproducibly, and efficiently can we measure left ventricular indices using M-mode, 2-dimensional, and 3-dimensional echocardiography in children?
- PMID: 18440346
- DOI: 10.1016/j.ahj.2007.11.034
How accurately, reproducibly, and efficiently can we measure left ventricular indices using M-mode, 2-dimensional, and 3-dimensional echocardiography in children?
Abstract
Background: Measurements of left ventricular (LV) size, mass, and function are the most common and important tasks for echocardiography in clinical practice and research in children with congenital and acquired heart diseases. There are little data to compare the utility of M-mode (MM), 2-dimensional (2D), and 3-dimensional (3D) echocardiographic techniques for quantification of LV indices. The objective of the study was to assess the accuracy, reproducibility, and efficiency of these echocardiographic methods for measurement of LV indices in children.
Methods: A prospective study was conducted in 20 consecutive children (mean 10.6 +/- 2.8 years, 11 male and 9 female subjects) using conventional MM, 2D, and real-time 3D echocardiography (RT3DE). A Sonos 7500 system (Philips Medical Systems, Andover, MA) was used. M-mode and 2DE measurements were made according to the American Society of echocardiography recommendations. To include the entire LV for volumetric measurement, full-volume 3D data sets were acquired from 4 electrocardiogram gated subvolumes. The 3DE measurements were made off-line manually using 4-plane and 8-plane algorithms by 4D Echo-View (TomTec Imaging Systems, Munich, Germany) and a semiautomated algorithm by QLAB (Philips Medical Systems). Magnetic resonance imaging studies were also performed to determine the LV indices by a disk summation method based on the Simpson principle.
Results: The correlation and agreement between MM, 2D, and RT3D echocardiography and magnetic resonance imaging measurements are good (r = 0.81-0.97) for the 3 methods. The correlation was superior for RT3DE compared with 2DE and MM. The correlation and agreement were similar for the three 3DE methods. The intra- and interobserver variabilities ranged from MM (4.3%-4.8% and 7.0%-8.7%), 2DE (3.3%-4.5% and 5.5%-7.3%), and 3DE (0.4%-2.3%, and 0.2%-4.8%). The total time (acquisition and analysis) used for MM measurements was the least compared with 2DE and 3DE. The total time for 3DE using the semiautomated algorithms was not significantly different compared with that for 2DE.
Conclusions: Our study showed that MM provides the most efficient assessment of LV indices but is the least accurate and reproducible technique compared with 2DE and 3DE. Three-dimensional echocardiography using both automated and manual analysis algorithm is superior to MM and 2DE for measurements of LV indices, and the automated 3DE algorithm is as efficient as 2DE. Therefore, 3DE using the automated algorithm is the method of choice for quantification of LV indices.
Similar articles
-
The Reliability of Automated Three-Dimensional Echocardiography-HeartModelA.I. Versus 2D Echocardiography Simpson Methods in Evaluation of Left Ventricle Volumes and Ejection Fraction in Patients With Left Ventricular Dysfunction.Med Arch. 2022 Aug;76(4):259-266. doi: 10.5455/medarh.2022.76.259-266. Med Arch. 2022. PMID: 36313951 Free PMC article.
-
Evaluation of Left Ventricular Volumes and Function by Real Time Three-Dimensional Echocardiography in Children with Functional Single Left Ventricle: A Comparison between QLAB and TomTec.Echocardiography. 2015 Oct;32(10):1554-63. doi: 10.1111/echo.12990. Epub 2015 Jun 22. Echocardiography. 2015. PMID: 26099327
-
Left ventricular volume measurement with echocardiography: a comparison of left ventricular opacification, three-dimensional echocardiography, or both with magnetic resonance imaging.Eur Heart J. 2009 Jan;30(1):98-106. doi: 10.1093/eurheartj/ehn484. Epub 2008 Nov 8. Eur Heart J. 2009. PMID: 18997179
-
Three-dimensional echocardiography to assess left ventricular geometry and function.Expert Rev Cardiovasc Ther. 2019 Nov;17(11):801-815. doi: 10.1080/14779072.2019.1697234. Expert Rev Cardiovasc Ther. 2019. PMID: 31770493 Review.
-
Real time three-dimensional echocardiography for quantification of ventricular volumes, mass, and function in children with congenital and acquired heart diseases.Echocardiography. 2013 Apr;30(4):472-82. doi: 10.1111/echo.12132. Echocardiography. 2013. PMID: 23551607 Review.
Cited by
-
Speckle tracking echocardiography detects decreased cardiac longitudinal function in anthracycline-exposed survivors of childhood cancer.Eur J Pediatr. 2016 Oct;175(10):1379-86. doi: 10.1007/s00431-016-2776-9. Epub 2016 Sep 13. Eur J Pediatr. 2016. PMID: 27620626
-
Imaging of patients with congenital heart disease.Nat Rev Cardiol. 2011 Nov 1;9(2):101-15. doi: 10.1038/nrcardio.2011.162. Nat Rev Cardiol. 2011. PMID: 22045045 Review.
-
Three-dimensional speckle tracking echocardiography for evaluation of ventricular function in patients with systemic lupus erythematosus: relationship between duration of lupus erythematosus and left ventricular dysfunction by using global longitudinal strain.Egypt Heart J. 2024 Jun 24;76(1):79. doi: 10.1186/s43044-024-00511-4. Egypt Heart J. 2024. PMID: 38914877 Free PMC article.
-
Left Ventricular Mass Quantification by Two-Dimensional Echocardiography in a Pediatric Population: Correlation with Cardiac Magnetic Resonance Imaging.Pediatr Cardiol. 2019 Feb;40(2):412-420. doi: 10.1007/s00246-018-1991-8. Epub 2018 Oct 11. Pediatr Cardiol. 2019. PMID: 30310940
-
Feasibility of Three-Dimensional Echocardiographic Analysis in Newborns by Novice Observers.Pediatr Cardiol. 2025 Aug;46(6):1694-1701. doi: 10.1007/s00246-024-03587-6. Epub 2024 Jul 17. Pediatr Cardiol. 2025. PMID: 39017728
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources