Comparision between bed side testing of blood glucose by glucometer vs centralized testing in a tertiary care hospital
- PMID: 18444586
Comparision between bed side testing of blood glucose by glucometer vs centralized testing in a tertiary care hospital
Abstract
Background: To determine the accuracy, turnaround time and cost effectiveness of bedside monitoring of blood glucose levels by non-laboratory health care workers and centralized testing of blood glucose by automated analyzer in a tertiary care hospital.
Methods: The study was conducted in Section of Chemical Pathology, Department of Pathology and Microbiology and Section of Endocrinology Department of Medicine, Aga Khan University and Hospital Karachi, from April 2005 to March 2006. One hundred and ten patients were included in the study. The blood glucose levels were analyzed on glucometer (Precision Abbott) by finger stick, using Biosensor Technology. At the same time venous blood was obtained to analyze glucose in clinical laboratory on automated analyzer (SYNCHRON CX7) by glucose oxidase method.
Results: We observed good correlation between bed side glucometer and laboratory automated analyzer for glucose values between 3.3 mmol/L (60 mg/dl) and 16.7 (300 mg/dl). A significant difference was observed for glucose values less than 3.3 mmol/L (p = 0.002) and glucose values more than 16.67 mmol/l (p = 0.049). Mean Turnaround time for glucometer and automated analyzer were 0.08 hours and 2.49 hours respectively. The cost of glucose testing with glucometer was 48.8% lower than centralized lab based testing.
Conclusion: Bedside glucometer testing, though less expensive does not have good accuracy in acutely ill patient with either very high or very low blood glucose levels.
Similar articles
-
Capillary versus venous bedside blood glucose estimations.Emerg Med J. 2005 Mar;22(3):177-9. doi: 10.1136/emj.2003.011619. Emerg Med J. 2005. PMID: 15735263 Free PMC article.
-
Evaluation of "point of care" devices in the measurement of low blood glucose in neonatal practice.Arch Dis Child Fetal Neonatal Ed. 2004 Jul;89(4):F356-9. doi: 10.1136/adc.2003.033548. Arch Dis Child Fetal Neonatal Ed. 2004. PMID: 15210675 Free PMC article.
-
Accuracy of bedside glucose measurement from three glucometers in critically ill patients.Crit Care Med. 2008 Nov;36(11):3062-6. doi: 10.1097/CCM.0b013e318186ffe6. Crit Care Med. 2008. PMID: 18824915
-
Robotics and automated workstations for rapid response testing.Am J Clin Pathol. 1995 Oct;104(4 Suppl 1):S26-32. Am J Clin Pathol. 1995. PMID: 7484946 Review.
-
Laboratory testing closer to the patient: the Canadian scene.Med Lab Sci. 1990 Oct;47(4):312-8. Med Lab Sci. 1990. PMID: 2283934 Review.
Cited by
-
Is serum zinc level associated with prediabetes and diabetes?: a cross-sectional study from Bangladesh.PLoS One. 2013 Apr 17;8(4):e61776. doi: 10.1371/journal.pone.0061776. Print 2013. PLoS One. 2013. PMID: 23613929 Free PMC article.
-
Social, organizational, and contextual characteristics of clinical decision support systems for intensive insulin therapy: a literature review and case study.Int J Med Inform. 2010 Jan;79(1):31-43. doi: 10.1016/j.ijmedinf.2009.09.004. Epub 2009 Oct 7. Int J Med Inform. 2010. PMID: 19815452 Free PMC article. Review.
-
Use of point-of-care glucometers during an oral glucose tolerance test in children for prediabetes and diabetes diagnosis: a comparison study.Adv Lab Med. 2023 Nov 29;5(2):189-196. doi: 10.1515/almed-2023-0089. eCollection 2024 Jun. Adv Lab Med. 2023. PMID: 38939205 Free PMC article.
-
Uso de glucómetros durante la prueba de tolerancia oral a la glucosa en niños para el diagnóstico de prediabetes y diabetes. Estudio comparativo.Adv Lab Med. 2024 Feb 19;5(2):197-204. doi: 10.1515/almed-2024-0017. eCollection 2024 Jun. Adv Lab Med. 2024. PMID: 38939199 Free PMC article. Spanish.
-
Blood glucose measurement in the intensive care unit: what is the best method?J Diabetes Sci Technol. 2013 Mar 1;7(2):489-99. doi: 10.1177/193229681300700226. J Diabetes Sci Technol. 2013. PMID: 23567008 Free PMC article. Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Other Literature Sources
Miscellaneous